one element: the resource handle to the chassis. Check your filter settings if the
resource handle only is not returned.

2.

Initialize the calibration process.
a.

Using the resource handle obtained with Find Hardware, set the 

CalPasswd

 property

to the current calibration password for the chassis.

Note

 

NI

 is the default calibration password.

b.

Using the same resource handle, set the 

CalExtAction

 property to 

OCXO Start

.

c.

Also using the same resource handle, call Save Changes to commit these changes to
hardware.

Note

 

To complete this step, LabVIEW must be running as an elevated

process with administrator rights. Otherwise, you will encounter error
-2147220611: 

The current user does not have permission

for the requested operation.

3.

Write a new value for the oscillator DAC value. The range of oscillator control voltage
values is 0 to 65535, with frequency increasing as the value increases. Use a DAC value
of 32768, which is in the middle of the valid range, as a starting point.
a.

Using the same resource handle, set the 

DacValue

 property to the new DAC value.

b.

Call Save Changes to commit these changes to hardware.

c.

Allow a wait time of 5 seconds after changing the DAC value.

4.

Measure the OCXO frequency following the steps in 

Verifying the OCXO Frequency

Accuracy

.

5.

Compare the measured frequency to the target frequency and repeat steps 3 and 4,
adjusting the oscillator DAC value up to increase the frequency and down to decrease the
frequency.

Note

 

NI recommends that you implement a binary search to determine the

optimal DAC value. This requires going through 16 iterations, with
confirmation that the final DAC value measured crosses over the target
frequency when the DAC value is changed by 1 bit.

6.

Compare the final measured frequency to the specified limits in Table 3 to confirm
adjustment completed successfully.

PXIe-1092 Calibration Procedure

  | 

© National Instruments

  | 

5

Summary of Contents for PXIe-1092

Page 1: ...following software on the calibration system PXI Platform Services 18 1 or later You can download PXI Platform Services from ni com downloads PXI Platform Services supports programming the calibration...

Page 2: ...beginning the calibration process The following setup and environmental conditions are required to ensure the PXIe 1092 chassis meets published specifications Keep cabling as short as possible Long c...

Page 3: ...Counter gate time to 3 seconds 4 Measure the OCXO frequency using the Frequency Counter 5 Compare the measured frequency to the device specifications To determine whether the device under test meets i...

Page 4: ...te You can use a binary search algorithm to find the optimal oscillator control voltage The PXIe 1092 uses a 16 bit digital to analog convertor DAC to create the oscillator control voltage therefore y...

Page 5: ...h frequency increasing as the value increases Use a DAC value of 32768 which is in the middle of the valid range as a starting point a Using the same resource handle set the DacValue property to the n...

Page 6: ...er adjustment is needed Note If any test fails reverification after performing an adjustment verify that you have met the Test Conditions before returning your device to NI Refer to Where to Go for Su...

Page 7: ...nts EULAs and third party legal notices in the readme file for your NI product Refer to the Export Compliance Information at ni com legal export compliance for the NI global trade compliance policy an...

Reviews: