User Manual 1178.8520.02 ─ 01
Useful commands for occupied bandwidth measurements described elsewhere
Configuring the channel:
●
[SENSe:]POWer:ACHannel:BANDwidth[:CHANnel<ch>]
●
●
[SENSe:]POWer:ACHannel:PRESet:RLEVel
Defining search limits:
●
CALCulate<n>:MARKer<m>:X:SLIMits[:STATe]
●
CALCulate<n>:MARKer<m>:X:SLIMits:LEFT
●
CALCulate<n>:MARKer<m>:X:SLIMits:RIGHt
Performing the measurement:
●
CALCulate<n>:MARKer<m>:FUNCtion:POWer<sb>:SELect
●
CALCulate<n>:MARKer<m>:FUNCtion:POWer<sb>[:STATe]
Retrieving results:
●
CALCulate<n>:MARKer<m>:FUNCtion:POWer<sb>:RESult?
Remote commands exclusive to occupied bandwidth measurements:
................................................................................................ 790
...........................................................................................790
[SENSe:]POWer:BWIDth
<Percentage>
[SENSe:]POWer:BANDwidth
<Percentage>
This command selects the percentage of the total power that defines the occupied
bandwidth.
Parameters:
<Percentage>
Range:
10 PCT to 99.9 PCT
*RST:
99 PCT
Default unit: PCT
Example:
POW:BAND 95PCT
Manual operation:
See
14.5.5.2
Programming Example: OBW Measurement
This programming example demonstrates the measurement example described in
Chapter 7.5.5, "Measurement Example"
on page 205 in a remote environment.
//-----------Preparing the measurement ------------
//Reset the instrument
*RST
//-------------Configuring the Measurement-----------------------
//Set the center frequency to 800 MHz.
FREQ:CENT 800MHz
//Set the reference level to -10 dBm.
Configuring and Performing Measurements