DISP:TRAC:Y:RLEV -10dBm
//Activate occupied bandwidth measurement.
CALC:MARK:FUNC:POW:SEL OBW
//Set the percentage of power to 99%.
POW:BWID 99PCT
//Set the channel bandwidth to 21 kHz.
POW:ACH:BAND 21kHz
//Optimize the instrument settings according to the channel bandwidth.
POW:ACH:PRES OBW
//Determine the ideal reference level for the measurement.
POW:ACH:PRES:RLEV
//Set the trace detector to positive peak.
DET APE
//--------------Performing the Measurement-----
//Select single sweep mode.INIT:CONT OFF
//Initiate a new measurement and waits until the sweep has finished.
INIT;*WAI
//---------------Retrieving Results-------------
//Return the occupied bandwidth.
CALC:MARK:FUNC:POW:RES? OBW
14.5.6
Measuring the Spectrum Emission Mask
All remote control commands specific to spectrum emission mask measurements are
described here.
See also
Chapter 14.5.2, "Configuring Power Measurements"
Managing Measurement Configurations
............................................................... 791
................................................................................793
Configuring a Multi-SEM Measurement
................................................................ 794
...................................................................................... 795
Configuring the Reference Range
.........................................................................808
.............................................................................810
Configuring MSR SEM Measurements
................................................................. 815
Configuring the List Evaluation
............................................................................. 821
........................................................................822
.................................................................................................822
............................................................................... 823
14.5.6.1
Managing Measurement Configurations
The following commands control measurement configurations for SEM measurements.
Configuring and Performing Measurements