Remote Commands
R&S
®
FSVA3000/ R&S
®
FSV3000
908
User Manual 1178.8520.02 ─ 01
14.7.4
Configuring Triggered and Gated Measurements
The commands required to configure a triggered or gated measurement in a remote
environment are described here.
The tasks for manual operation are described in
Chapter 8.6, "Trigger and Gate Con-
The commands required for trigger input or output are described in
"Configuring the Trigger Output"
*OPC should be used after requesting data. This will hold off any subsequent changes
to the selected trigger source, until after the sweep is completed and the data is
returned.
Configuring the Triggering Conditions
...................................................................908
Configuring Gated Measurements
........................................................................ 913
Configuring the Trigger Output
..............................................................................916
14.7.4.1
Configuring the Triggering Conditions
The following commands are required to configure a triggered measurement.
...........................................................................................908
TRIGger[:SEQuence]:HOLDoff[:TIME]
..............................................................................908
TRIGger[:SEQuence]:IFPower:HOLDoff
........................................................................... 909
TRIGger[:SEQuence]:IFPower:HYSTeresis
.......................................................................909
TRIGger<tp>[:SEQuence]:LEVel[:EXTernal<port>]
............................................................ 910
TRIGger[:SEQuence]:LEVel:IFPower
............................................................................... 910
TRIGger[:SEQuence]:LEVel:IQPower
...............................................................................910
TRIGger[:SEQuence]:LEVel:VIDeo
.................................................................................. 911
.......................................................................................... 911
........................................................................................ 911
TRIGger[:SEQuence]:TIME:RINTerval
..............................................................................912
TRIGger[:SEQuence]:DTIMe
<DropoutTime>
Defines the time the input signal must stay below the trigger level before a trigger is
detected again.
Parameters:
<DropoutTime>
Dropout time of the trigger.
Range:
0 s to 10.0 s
*RST:
0 s
Default unit: S
Manual operation:
See
TRIGger[:SEQuence]:HOLDoff[:TIME]
<Offset>
Defines the time offset between the trigger event and the start of the sweep.