Other Commands:

0xA0: Firmware Version Query 

The firmware version information will be sent and data will be 
terminated with 0x00 character.

0xA1: Battery Voltage Query (mV)

Battery voltage information will be sent and data will be 
terminated with 0x00 character.

0xA2: Close Sleep Timer (10 seconds)

If there is no connection with the Bluetooth probe for more than 
10 seconds, the port goes into sleep mode and power saving 
mode will be on. When the Bluetooth connection is turned on, the 
sleep mode will be off and the port will work as usual and user 
operation is not affected by this process. 
Confirmation information will be sent and data will be terminated 
with 0x00 character.

0xC2: Shut Down The Probe

Using this command, the power can be turned off as if the optical 
probe was closed by the power button.

0xC3: Auto Shutdown Timer Setting

A time interval between 0 to 254 seconds should be set (in HEX 
format). If there is no communication with the probe after that 
time interval, the device will be turned off (probe shuts down). 
This can prevent the probes to be left forgotten with power on for 
long time and consume battery. 
The default value is 254 seconds. This feature can be disabled by 
setting value 255.
After sending the command, confirmation information will be sent 
and data will be terminated with character 0x00. Then the time 
interval must be set. The next confirmation information will be 
terminated with character 0x00 again.

0xD0: Probe Reading Sensitivity Setting

If user gets some meaningless or corrupted data while reading 
the meter, it is possible to change the reading accuracy of the 
probe and make special setting based on meter type to prevent 
reflections or light interferences.

A value between 0 to 255 should be set (in HEX format). As this 
value increases, sensitivity of the probe decreases and light 
reception capacity falls. The default value is 49 (0x31).
After  sending  the  command,  confirmation  information  will  be 
sent and data will be terminated with character 0x00. Then the 
sensitivity value must be set. The next confirmation information 
will be terminated with character 0x00 again.
The user can transmit any data and command to the optical 
probe without any restrictions and can use the optical probe 
at any desired data type and communication speed at any time. 
The commands transmitted to the optical probe and the data 
transmitted to the meter is always separated.

0xE0: Save Settings (available with v3 designs and later)

It is possible to save settings for following data Sleep Timer, Auto Power 
seconds, Reading Sensitivity, Baud Rate and Data type. After sending 
the command, confirmation information will be sent and data will be 
terminated with character 0x00. All listed values are saved as it is set and 
next time the device restarted they will be used based on saved values.

0xE1: Restore Factory Defaults ( available with v3 designs and later)

Command sets values as follows and saves them

•  Sleep Timer: ON
•  Auto Power Down Seconds : 254
•  Reading Sensitivity: 49
•  Baud Rate: 300
•  Data Type: 7E1

After sending the command, confirmation information will be sent 
and data will be terminated with character 0x00.