When operating in Auto IP mode, the instrument automatically acquires the IP
address from 169.254.0.1 to 169.254.255.254 and subnet mask 255.255.0.0
according to the current network configuration.
•
When all the three configuration modes are set to "On", the priority of
parameter configuration is "DHCP", "Auto IP", and "Manual IP". Therefore, to
use the Auto IP configuration mode, “DHCP
(
:SYSTem:COMMunicate:LAN:DHCP[:STATe]
)” should be set to "OFF".
•
The three IP configuration modes cannot be set to "OFF" at the same time.
•
After sending the command, execute
apply the network parameters currently set.
Return Format
The query returns 1 or 0.
Examples
:SYST:COMM:LAN:AUTO ON /*Enables the Auto IP mode.*/
:SYST:COMM:LAN:AUTO? /*Queries the on/off state of Auto IP. The
query returns 1.*/
4.11.4.3
:SYSTem:COMMunicate:LAN:DHCP[:STATe]
Syntax
:SYSTem:COMMunicate:LAN:DHCP[:STATe]
<
bool>
:SYSTem:COMMunicate:LAN:DHCP[:STATe]?
Description
Sets or queries the on/off state of DHCP.
Parameter
Name
Type
Range
Default
<bool>
Bool
{0|1|ON|OFF}
ON|1
Remarks
•
In DHCP mode, the DHCP server in the current network assigns network
parameters (such as the IP address) to the instrument.
Command System