Command Systems
RIGOL
Programming Guide for DS1000E, DS1000D Series
2-27
PULSe Trigger
1.
:TRIGger:PULSe:MODE
Command Format:
:TRIGger:PULSe:MODE
<mode>
:TRIGger:PULSe:MODE?
Function Explanation:
The commands set and query the pulse condition. <mode> could be
+GREaterthan (positive pulse greater than), +LESSthan (positive pulse less
than), +EQUal (positive pulse equals to), -GREaterthan (negative pulse greater
than), -LESSthan (negative pulse less than) or –EQUal (negative pulse equals
to).
Returned Format:
The query rGREATER THAN, +LESS THAN, +EQUAL, -GREATER THAN,
-LESS THAN or -EQUAL.
Example:
:TRIG:PULS:MODE
+GRE Setup the trigger condition as +GREaterthan.
:TRIG:PULS:MODE?
The query rGREATER THAN.
2.
:TRIGger:PULSe:SENSitivity
Command Format:
:TRIGger:PULSe:SENSitivity <count>
:TRIGger:PULSe:SENSitivity?
Function Explanation:
The commands set and query the sensitive of pulse trigger. The range of
<count> could be 0.1div~1div.
Returned Format:
The query returns the setting value of <count> in div.
Example: