50
How can I...
Set the generator signal symmetry
word SetFuncGenSymmetry( double dSymmetry )
Description:
This routine sets the symmetry of the output signal. The symme-
try can be set between 0 and 100. With a symmetry of 50, the
positive part of the output signal and negative part of the output
signal are equally long. With a symmetry of 25, the positive part
of the output signal takes 25% of the total period and the negati-
ve part takes 75% of the total period.
With signal types
DC
,
Noise
and
Arbitrary
, the symmetry value is
ignored.
Input:
dSymmetry
The requested symmetry value:
0 <= value <= 100
Output:
Return value
E_NO_ERRORS
E_NO_GENERATOR
E_INVALID_VALUE
E_NO_HARDWARE
Get the current generator signal symmetry
word GetFuncGenSymmetry( double *dSymmetry )
Description:
This routine retrieves the currently selected symmetry of the
output signal.
Input:
-
Output:
dSymmetry
the current symmetry value
Return value
E_NO_ERRORS
E_NO_GENERATOR
E_INVALID_VALUE
E_NO_HARDWARE