SCPI Command Reference
Light Source
63
Product-Specific Commands—Description
:SOURce[1..n]:AM:INTernal:
FREQuency?
Description
This query returns a value indicating the current
internal modulation frequency. If the source is in
CW mode, the function will return 0.
*RST sets the modulation frequency to 0 Hz
(CW).
Syntax
:SOURce[1..n]:AM:INTernal:FREQuency?
Parameter(s)
None
Response Syntax
<ModulationFrequency>
Response(s)
ModulationFrequency:
The response data syntax for
<ModulationFrequency> is defined as a <NR3
NUMERIC RESPONSE DATA> element.
The <ModulationFrequency> response is the
internal modulation frequency of the source, in
Hz. If the source is in CW mode, the returned
value is 0.
Example(s)
SOUR:POW:STAT ON
SOUR:AM:INT:FREQ 270
SOUR:AM:INT:FREQ?
See Also
SOURce[1..n]:AM:INTernal:FREQuency