98
PR-805/810/810
L
User Manual
Setup
Command
Description
SW
Internal Neutral Density Filter Select
Open/Enable internal ND filter, filter attenuation are open, 10X, 100X, 1000X, Auto
Syntax
: SWn[CR]
where
: n = Enable/ Open ND
0 = Open (No ND)
1 = ND 1 (10X attenuation)
2 = ND 2 (100X attenuation)
3 = ND 3 (1000X attenuation)
99 = Auto ND
Response
: 0000[CR][LF] If all OK, else
NNNN[CR][LF] (NNNN = Error code)
SZ
Measure Shutter Control
Controls the actions of the measure shutter. When set to 0 the measure shutter will be
closed after each measurement and a dark measurement will be taken. When Measure
shutter control is set to 1, the measure shutter will never close and no dark measurements
will be taken.
It is recommended that measurements be taken with Measure shutter
control set to 0 so that a dark measurement can be captured
.
Syntax
: SZs[CR]
Where:
s = Shutter Control
0 = Close after a measurement
1 = Never close (always open)
Response
: 0000[CR][LF] If all OK, else
NNNN[CR][LF] (NNNN = Error code)