5-7
SeCTIon 5
Using Substrate Bias
FET _ Comm _ Source _ Vsb(1, 12, 10, 1, 12,
80, –2)
to the instrument .
The sources will be enabled, and the substrate bias is applied,
10 .
the gate-source voltage value is applied, and the drain-source
sweep is executed . The gate-source voltage value is then incre-
mented and the drain-source sweep is re-run .
Once the gate-source sweep has been completed, the data
11 .
(V
SB
, V
GS
, V
DS
, and I
D
) will be presented in the Instrument
Console window of TSB .
5.3.8 Typical Program 13 Results
Figure 5-6
shows a typical plot generated by Example
Program 13 .
5.3.9 Program 13 Description
Both instruments are returned to default conditions . Node 1
SMUB, which sweeps V
GS
, is configured as follows:
Source V
•
1mA compliance, autorange
•
Local sense
•
vgsstart
•
: 0V
vgsstop
•
: 10V
vgssteps
•
: 5
Next, Node 1 SMUA, which sweeps V
DS
and measures I
D
, is set up
to operate in the following manner:
Source V
•
Local sensing
•
100mA compliance, autorange measure
•
1 NPLC Line cycle integration
•
vdsstart
•
: 0V
vdsstop
•
: 10V
vdssteps
•
: 100
Finally, Node 2 SMUA, which provides substrate bias, is pro-
grammed as follows:
Source V
•
Local sensing
•
10mA compliance, autorange measure
•
Both instruments are returned to default conditions; the sources
are zeroed and enabled . The substrate bias (V
SB
) and gate-source
(V
GS
) are applied and the program enters the main program loop
to perform five I
D
vs . V
DS
sweeps, one for each of five V
GS
values .
Node 1 SMUA then cycles through its sweep list, setting V
DS
to the
required values, and measuring I
D
at each step along the way . The
program then loops back for the next sweep until all five sweeps
have been performed .
Next, all three SMU outputs are zeroed and disabled . Finally, the
data is written to the Instrument Console of the TSB .
5.3.10 Modifying Program 13
For different sweeps, the V
GS
start, V
GS
stop, V
GS
steps, V
DS
start,
V
DS
stop, and V
DS
steps values can be changed as required . For
different sweep lengths, array size and loop counter values must
be adjusted accordingly .
5.4 BJT Substrate Biasing
The following paragraphs discuss using one dual-channel and
one single-channel Series 2600 System SourceMeter instrument
to perform tests on a four-terminal device, such as a BJT, with
substrate bias . The example shown in this section is a modified
version of the common-emitter BJT test presented previously in
the guide .
5.4.1 Program 14 Test Configuration
Figure 5-7
shows the test configuration for Program 14 . Node 1
SMUB is used to sweep I
B
, while Node 1 SMUA sweeps V
CE
and
measures I
C
. Node 2 SMUA applies the substrate bias (V
SB
) to the
device under test .
5.4.2 example Program 14: Common-
emitter Characteristics with a
Substrate Bias
Program 14 demonstrates common-emitter characteristic test pro-
gramming with substrate bias . Proceed as follows:
With the power off, connect the dual-channel System Source-
1 .
Meter instrument to the computer’s IEEE-488 interface . Con-
nect the single-channel System SourceMeter instrument to the
dual-channel master using a crossover Ethernet cable .
Connect the test fixture to both units using appropriate
2 .
cables .
Turn on the instruments and allow the units to warm up for
3 .
two hours for rated accuracy .
Configure the TSP-Link communications for each instrument .
4 .
Slave:
A single-channel instrument such as the Model 2601,
2611, or 2635 .
1 . Press the MENU key to access MAIN MENU .
Summary of Contents for Series 2600
Page 5: ......
Page 7: ......
Page 17: ...2 8 Section 2 Two terminal Device Tests...
Page 39: ...4 10 Section 4 FET Tests...
Page 53: ...6 4 Section 6 High Power Tests...
Page 130: ...A 77 Appendix A Scripts BJT_Comm_Emit_Vsb...
Page 136: ...Click below to find more Mipaper at www lcis com tw Mipaper at www lcis com tw...