Models 2510 and 2510-AT
Calibration Reference
B-7
:TEMPerature
(:CALibration:PROTected:SENSe:TEMPerature)
(:CALibration:PROTected:SENSe:TEMPerature:DATA?)
Purpose
To calibrate the temperature sensor measurements.
Format
:cal:prot:sens:temp <Value>
Parameters
<Resistor_Value> = 100k
Ω
thermistor: 0 to 125e3[
Ω
]
10k
Ω
thermistor: 0 to 12.5e3[
Ω
]
1k
Ω
thermistor: 0 to 1.25e3[
Ω
]
100
Ω
thermistor: 0 to 125[
Ω
]
100
Ω
RTD: 0 to 125[
Ω
]
1000¾ RTD: 0 to 1250 [¾]
VSS: 0V to 5[V]
ISS: 350 to 1.25e3[
Ω
] then 0.01 to 1.5[V]
Query
:cal:prot:sens:temp:data?
Description
The :CAL:PROT:SENS:TEMP command calibrates the Model 2510 tem-
perature sensor measurements. During the calibration process, a 100
Ω
, 1k
Ω
,
10k
Ω
, or 100k
Ω
(nominal) resistor or a short must be connected to the
INPUT terminals, and the characterized value, a zero value, or a voltage
reading is sent as the command parameter, depending on the calibration step
(see Section 2 for details). The appropriate sensor and range must first be
selected with the following commands:
•
Enable 4-wire sensing:
:SYST:RSEN ON
•
Thermistor sensor:
:SENS:TEMP:TRAN THER
•
Thermistor range:
:SENS:TEMP:THER:RANG <Range>
•
RTD sensor:
:SENS:TEMP:TRAN RTD
•
RTD range:
:SENS:TEMP:RTD:RANG <Range>
•
ISS sensor:
:SENS:TEMP:TRAN ISS
•
VSS sensor:
:SENS:TEMP:TRAN VSS
The :CAL:PROT:SENS:TEMP:DATA? query allows you request the tem-
perature sensor calibration parameters.
Example
:CAL:PROT:SENS:TEMP 100.5
Calibrate temperature sensor.
2510-902-01 (E - Mar 2010)(Service).book Page 7 Tuesday, March 16, 2010 1:04 PM
Summary of Contents for TEC SourceMeter 2510
Page 3: ...2510 902 01 E Mar 2010 Service book Page 2 Tuesday March 16 2010 1 04 PM...
Page 28: ...2 Calibration 2510 902 01 E Mar 2010 Service book Page 1 Tuesday March 16 2010 1 04 PM...
Page 54: ...3 RoutineMaintenance 2510 902 01 E Mar 2010 Service book Page 1 Tuesday March 16 2010 1 04 PM...
Page 58: ...4 Troubleshooting 2510 902 01 E Mar 2010 Service book Page 1 Tuesday March 16 2010 1 04 PM...
Page 76: ...5 Disassembly 2510 902 01 E Mar 2010 Service book Page 1 Tuesday March 16 2010 1 04 PM...
Page 82: ...6 ReplaceableParts 2510 902 01 E Mar 2010 Service book Page 1 Tuesday March 16 2010 1 04 PM...
Page 92: ...A Specifications 2510 902 01 E Mar 2010 Service book Page 1 Tuesday March 16 2010 2 33 PM...
Page 110: ...C CalibrationProgram 2510 902 01 E Mar 2010 Service book Page 1 Tuesday March 16 2010 1 04 PM...
Page 119: ...2510 902 01 E Mar 2010 Service book Page 4 Tuesday March 16 2010 1 04 PM...