80
Keysight E1460A 64-Channel Relay Multiplexer User Guide
Relay Multiplexer Command Reference
OUTPut
“Trig Out” Port Shared by Switchboxes:
Only one switchbox configuration can use the
selected trigger at a time. When enabled, the “Trig Out” port is pulsed by the
switchbox each time a scanned channel is closed. To disable the output for a
specific switchbox, send the OUTP OFF or 0 command for that switchbox.
One Output Selected at a Time:
Only one output (ECLTrg0 or 1; TTLTrg0, 1, 2, 3, 4, 5, 6,
or 7; or EXTernal) can be enabled at one time. Enabling a different output source
will automatically disable the active output.
For example, if TTLTrg1 is the active output and TTLTrg4 is enabled, TTLTrg1 will
become disabled and TTLTrg4 will become the active output.
Related Commands:
[ROUTe:]SCAN, TRIGger:SOURce, OUTPut[:EXTernal][:STATe]?
*RST Condition:
OUTPut[:EXTernal][:STATe] OFF (disabled)
Example
Enabling "Trig Out" Port
OUTP:EXT 1 !Enable "Trig Out" port to output pulse
after each scanned channel is closed
OUTPut[:EXTernal][:STATe]?
OUTPut[:EXTernal][:STATe]?
queries the present state of the "Trig Out" port. The
command returns "1" if the port is enabled or "0" if disabled.
Example
Query "Trig Out" Port Enable State
This example enables the "Trig Out" port and queries the enable state. The
OUTPut? command returns "1" since the port is enabled.
OUTP:EXT ON !Enable “Trig Out” port
OUTP:EXT? !Query port enable state
OUTPut:TTLTrg
n
[:STATe]
OUTPut:TTLTrg
n
[:STATe]
<mode>
selects and enables which TTL Trigger bus line (0 to
7) will output a trigger when a channel is closed during a scan. This is also used
to disable a selected TTL Trigger bus line. "
n
" specifies the TTL Trigger bus line (0
to 7) and "
mode
" enables (ON or 1) or disables (OFF or 0) the specified TTL
Trigger bus line.
Parameters
Name
Type
Range of Values
Default
<
n
>
numeric
0 or 7
N/A
<
mode
>
boolean
0 I 1 I OFF I ON
OFF I 0