5

 

.

 

 

Serial Communication 

 

 

 

 

 

 

 

Before using the USB communication port, download the USB driver from http://www.measuretek.net/Download.html 

 
Detailed Serial (USB) Output Information: 

 

U5=0: No serial communication. No data transmitted or received even if the scale is installed with serial communication 

hardware.  

 

U5=1: Pressing PRINT/HOLD will output the current stable weight, weight unit, and current status data in the following format, 

but will not receive data: 

<LF>< weight, minus, decimal point, weight unit><CR><LF> H1H2H3 <CR><ETX> 

 

U5=2: Pressing PRINT/HOLD will output the current stable gross weight, tare, net weight, weight unit and current status data in 

the following format, but will not receive data: 

 

<LF><Gross: reading, minus, decimal point, unit><CR> 

<LF> <Tare: reading, decimal point, unit><CR> 

<LF> <Net: reading, minus, decimal point, unit><CR> 

<LF> H1H2H3<CR><ETX> 

    Bytes used: 

8  -  Weight reading 

1  -  Minus 

1  -  Decimal point 

2 or 5  -    Weight unit 

3  -  Current status (H1.H2.H3) 

 

 

U5=3: Continuously outputs the current displayed reading, weight unit, and current status data, but will not receive data.    The 

output format is the same as U5=1.   

 

U5=4: Continuously outputs the current gross weight, tare weight, net weight data, weight unit, and current status data, but will 

not receive data. The output format is the same as U5=2. 

 

U5=5: When the scale is stable, it will output the current displayed reading ,weight unit, and current status data automatically 

one time, but will not receive data. The output format is same as U5=1. 

 

U5=6: When the scale is stable, it will output the current gross weight, tare weight, net weight unit, and current status data 

automatically one time, but will not receive data. The output format is same as U5=2.   

 

U5=7: Two way communication.    Responses to serial commands will be immediate, or within one   

Power 

Loadcell

USB 

Summary of Contents for PS-102-200

Page 1: ...W W W M E A S U R E T E K N E T PS 102 Operation Manual Contents subject to change without notice Version 1 0 11 2011 ...

Page 2: ......

Page 3: ...vices that emit radio frequencies that may cause unstable readings If the scale performs poorly try moving to a different room or location Avoid using in heavy vibration or heavy airflow conditions Specifications Model PS 102 75 PS 102 200 Max Capacity 165lb 75kg 440lb 200kg Readability 0 1lb 0 05kg 0 2lb 0 1kg Resolution 1 1500 1 2000 Construction Epoxy painted carbon steel Stainless steel platfo...

Page 4: ...ition from right to left UNIT Weighing mode Chooses weight unit Setup Calibration mode Increases the digit in the flashing data entry position by 1 TARE Weighing mode Tares the weight Setup or Calibration mode Confirms the input data and forwards to next step ON OFF ZERO Weighing mode 3 seconds Zeros the platform weight 3 seconds Powers off the scale Setup or Calibration mode Exits to normal weigh...

Page 5: ... The default calibration input weight will be 50 of the scale s rated capacity Use the PRINT HOLD and UNIT keys adjust to the standard calibration weight you will use Note do not use less than 10 of the scale s rated capacity Place the standard calibration weight on the platform and press TARE to confirm your input and move to the next step If CAL Er is displayed press ZERO ON OFF key to exit this...

Page 6: ...ight within 3 seconds Note When weight exceeds 10d and the variation is within 2d 50d the stable weight is displayed until less than 10d is detected on the platform 1d 1 division 1 unit of readability as noted in Specifications U5 1 Serial communication output format 0 Communication is disabled 1 Outputs stable weight unit and status when PRINT HOLD is pressed but does not receive data 2 Output gr...

Page 7: ...et reading minus decimal point unit CR LF H1H2H3 CR ETX Bytes used 8 Weight reading 1 Minus 1 Decimal point 2 or 5 Weight unit 3 Current status H1 H2 H3 U5 3 Continuously outputs the current displayed reading weight unit and current status data but will not receive data The output format is the same as U5 1 U5 4 Continuously outputs the current gross weight tare weight net weight data weight unit ...

Page 8: ...haracter indicates positive weight The minus sign follows after the first digit Leading zero digits are suppressed Serial Communication Symbols LF Line Feed character hex 0AH CR Carriage Return character hex 0DH ETX End of Text character hex 03 SP Space hex 20H H1H2H3 Three status bytes see below for byte definitions p Polarity minus sign for negative weight and space for positive weight W1 W8 Wei...

Page 9: ...F H1H2H3 CR ETX Zero function is activated and returns to current scale status like pressing ZERO ON OFF If zero function cannot be activated it will return to current scale status Command T CR 54h 0dh Response LF H1H2H3 CR ETX Tare function is activated and returns to current scale status like pressing TARE If tare function cannot be activated it will return to current scale status Command U CR 5...

Page 10: ... 3 6V Replacement Parts Part Number Description MH12R97601G AC120V DC6V 500mA UL adaptor One Year Limited Warranty MeasureTek products covered in this manual are guaranteed to be free from defects in material and workmanship for a period of one year after date of purchase Misuse accidental damage overload alteration and improper installation are expressly excluded Any product which is determined t...

Reviews: