PS
‐
102
www.measuretek.net
.
7
.
Serial Communication (continued)
Serial Commands and Responses
Command:
S<CR> (53h 0dh)
Response:
<LF> H1H2H3<CR><ETX>
Command:
W<CR> (57h 0dh)
Responses:
over capacity:
<LF>^^^^^^^^^^u1u2<CR><LF>H1H2H3<CR><ETX>
under capacity:
<LF>_________u1u2<CR><LF> H1H2H3<CR><ETX>
zero-point error:
<LF>----------u1u2<CR><LF> H1H2H3<CR><ETX>
Note: If the weight unit is lb:oz, U1U2 = “lb oz” in above item
①②③
.
Normal weight is displayed, current weight unit is kg or lb:
<LF><p>w1w2w3w4w5w6<dp>w7w8u1u2<CR><LF>H1H2H3<CR><ETX>
Normal weight is displayed, current weight unit is lb:oz,
<LF><p>w1w2w3w4w5w6lb<sp>w7w8<o><z><CR>H1H2H3<CR><ETX>
or
<LF><p>w1w2w3w4w5lb<sp> w6w7<dp>w8oz<CR>H1H2H3<CR><ETX>
Command:
Z<CR> (5ah 0dh)
Response:
<LF> H1H2H3<CR><ETX>
Zero function is activated and returns to current scale status, like pressing ZERO/ON/OFF. If zero function cannot be
activated, it will return to current scale status.
Command:
T<CR> (54h 0dh)
Response:
<LF> H1H2H3<CR><ETX>
Tare function is activated and returns to current scale status, like pressing TARE . If tare
function cannot be activated, it will return to current scale status.
Command:
U<CR> (55h 0dh)
Response:
<LF>u1u2<CR><LF> H1H2H3<CR><ETX>
Changes unit of measure and returns to current scale status with new units, like pressing UNIT button.
Command
: L<CR> (4ch 0dh)
Response:
<LF> H1H2H3<CR><ETX>
If hold function is enabled, goes to or exits from hold mode.
Command:
X<CR> (58h 0dh)
Response: none
Powers off the scale, like pressing ZERO/ON/OFF.
Command: All others
Response:
<LF>? <CR><ETX>
Unrecognized command.
Load Cell Wiring
PIN1 ------ Exci
PIN2 ------
PIN3 ------ Signal -
PIN4 ------ Excitation –
Summary of Contents for PS-102-200