NI 5105 Calibration Procedure
3.
Configure the common vertical properties using the niScope Configure Vertical VI.
4.
Configure the horizontal properties using the niScope Configure Horizontal Timing VI.
5.
Commit all the parameter settings to hardware using the niScope Commit VI.
LabVIEW VI
C/C++ Function Call
Call
niScope_ConfigureVerti
cal
with the following
parameters:
coupling
:
NISCOPE_VAL_DC
probeAttenuation
:
1.0
vi
: The instrument handle from
niScope_init
channelList
: The name of the
channel being tested
range
: The
Range
value listed in
Table 6 for the current iteration
offset
:
0.0
enabled
:
NISCOPE_VAL_TRUE
LabVIEW VI
C/C++ Function Call
Call
niScope_Configure
HorizontalTiming
with
the following parameters:
enforceRealtime
:
NISCOPE_VAL_TRUE
numRecords
:
1
vi
: The instrument handle
from
niScope_init
minSampleRate
:
10,000,000
refPosition
:
50.0
minNumPts
:
100,000
LabVIEW VI
C/C++ Function Call
Call
niScope_Commit
with
the following parameter:
vi
: The instrument handle from
niScope_init