Omega Model CYD211 User’s Manual
4-16 Remote
Operation
ALARM?
Input Alarm Parameter Query
Input:
ALARM?[term]
Returned:
<off/on>, <high value>, <low value>, <deadband>, <latch
enable> [term]
Format:
n,+nnn.n,+nnn.n,+nn.n,n
(Refer to command for description)
ALMRST
Reset Alarm Status Command
Input:
ALMRST[term]
Remarks:
Clears both the high and low status of the alarm, including
latching alarm.
ANALOG
Analog Output Parameter Command
Input:
ANALOG <mode>, <range>[term]
Format:
n,n
<mode> Specifies mode in which analog output operates
where 0 = voltage mode and 1 = current mode.
<range> Sets temperature range that analog output uses as
full scale.
0 = 0 – 20 K
3 = 0 – 325 K
1 = 0 – 100 K
4 = 0 – 475 K
2 = 0 – 200 K
5 = 0 – 1000 K
Example: ANALOG
0,1[term]
– Sets analog output to voltage mode
(0
–
10V) 100.0 K at +100% output (+10.0 V) and 0.0 K at
0% output (0.0 V).
ANALOG?
Analog Output Parameter Query
Input:
ANALOG?[term]
Returned:
<mode>, <range>
[term]
Format:
n,n
(Refer to command for definition)
AOUT?
Analog Output Data Query
Input:
AOUT?[term]
Returned:
<analog output>
[term]
Format:
+nnn.nn
Remarks:
Returns the percentage of output of the analog output.