14
1051_2_Product_Manual - May 5, 2010 4:34 PM
Mechanical Drawing
1:1 scale
Device Specifications
Characteristic
Value
Electrical
USB-Power Current Specification
100mA
Device Current Consumption
25mA
USB Device Voltage
4.75 - 5.25 VDC
Accuracy
Temperature Update Rate
25 samples/second
Ambient Temperature Resolution
±0.5°C
Thermocouple Resolution (°C)
0.1°C (K-type @ 25°C)
Thermocouple Resolution (
µ
V)
4
µ
V
Thermocouple Accuracy
2°C (K-type)
Typical Noise
0.02°C standard deviation
Typical Thermocouple Temperature Range
E-Type
-200°C to +900°C
J-Type
0°C to +750°C
K-Type
-200°C to +1250°C
T-Type
-200°C to +350°C