Remote Commands
R&S
®
FSVA3000/ R&S
®
FSV3000
1040
User Manual 1178.8520.02 ─ 01
//------------- Configuring the signal count marker -------------------
CALC:MARK1 ON
//Activates marker1
CALC:MARK1:COUN ON
//Switches on the frequency counter for marker 1.
CALC:MARK1:COUN:RES 1kHz
//Sets the resolution of the frequency counter to 1kHz
//------- Performing the measurement -------------------------
INIT;*WAI
//Performs a measurement and waits for it to end
//---------------Retrieving Results-------------
CALC:MARK1:COUN:FREQ?
//Returns the signal counter value as the precise marker frequency.
14.8.4
Configuring Display and Limit Lines
The commands required to configure display and limit lines in a remote environment
are described here.
................................................................................... 1040
......................................................................................... 1043
14.8.4.1
Configuring Display Lines
The following commands configure vertical and horizontal display lines.
.............................................................................................1040
.................................................................................. 1041
............................................................................................. 1041
...................................................................................1041
............................................................................................. 1042
...................................................................................1042
CALCulate<n>:DLINe<dl>
<Position>
This command defines the (horizontal) position of a display line.
Suffix:
<n>
.
<dl>
1 | 2
Parameters:
<Position>
The value range is variable.
You can use any unit you want, the R&S
FSV/A then converts
the unit to the currently selected unit. If you omit a unit, the
R&S
FSV/A uses the currently selected unit.
*RST:
(state is OFF)
Default unit: DBM
Analyzing Measurements (Basics)