Serial
Communications
PT12
Manual
7/20/2021
Page
42
of
48
Channel
Commands
Typical
Response
(Read
Command)
or
Input
Values
(Write
Command)
Notes
T1?
25.12345
Temperature
in
°C
for
channel
1
T1.I?
T1.I=
0
Probe’s
current
mode.
0=0.33
mA
1=0.66
mA
2=1
mA
Will
be
ignored
if
AUTO_CURRENT_DISABLE
is
not
1.
T1.AUTO_CURRENT_DISABLE?
T1.AUTO_CURRENT_DISABLE=
0
1=Auto
current
mode
disabled
0=Auto
current
mode
enabled
T1.CONNECTED?
1
0=
Probe
Disconnected
1=
Probe
Connected
T1.OHMS?
99.123
Resistance
of
probe
in
ohms
T1.AVGCOUNT?
T1.AVGCOUNT=
8
Number
of
measurements
to
include
in
average
of
value.
Typical
range
0
‐
100
T1.SAVE=RHS
N/A
SAVE=RHS
to
save
the
config.
RHS
must
be
uppercase.
T1.DEFAULT=RHS
N/A
DEFAULT=RHS
to
set
the
channel
config
to
default.
Does
not
save
without
running
a
save
command.
RHS
must
be
uppercase.
Note:
All
of
the
channel
commands
for
T2
‐
T12
are
identical
to
the
commands
above.
Just
replace
T1
with
T2
‐
T12
for
the
desired
channel.