3.4 Commands
3 – 15
•
SCAE
0.0
•
SCAT
0.1
•
SCAS
ONEDIR
•
SCAC
ONCE
•
SCAD
ON
•
SCAA
OFF
•
KCLK
ON
•
ALRM
ON
The same reset to default configuration can also be performed by
holding the INPUT [Enable] button depressed while switching on
power.
3.4.9
Status commands
Interlock
ILOC?
Reads the current value of the interlock. If the safety interlock is
asserted,
ILOC?
returns
1
; otherwise it returns
0
.
ILOC?
Example:
1
Overload
OVLD?
Reads the current value of the signal overload status. Returns
1
if
the output is overloaded (current limit mode), or
0
otherwise.
OVLD?
Example:
0
Status byte
*STB? [
i
]
Reads the Status Byte register [bit
i
].
*STB?
Example:
0
Service request enable
*SRE(?) [
i,
]
{
j
}
Set (query) the Service Request Enable register [bit
i
]
{
to
j
}
.
*SRE 0,1
Example:
DC205 Precision DC Voltage Source