©2022 Televac® - The Fredericks Company | www.televac.com | sales@televac.com | +1 215 947 2500
mx7b_im rev E Page
19
of
23
9.4.3
RS-485 and USB General Write Commands
Command
Response
Description
W1xx
xx
Set units, where xx is:
PA = Pa
TR = Torr
MB = mbar
W3
xx
Toggle resolution, where xx is:
LO = low
HI = high
W51yzz
1yzz
Set analog output format to linear:
yzz = high value
y = high value exponent sign
zz = high value exponent
W52yzzabb
2yzzabb
Set analog output format to logarithmic:
yzz = high value
abb = low value
y = high value exponent sign
zz = high value exponent
a = low value exponent sign
bb = low value exponent
W53
3
Set analog output format to linear-by-decade
W6xx
xx
Set gas type, where xx is:
N2 = Nitrogen
AR = Argon
H2 = Hydrogen
HE = Helium
NE = Neon
KR = Krypton
C2 = Carbon Dioxide
Response is new gas type.
W7YppsePPSE
ppseePPSEE
Set thresholds for set point Y, where:
ppsee = SPL
PPSEE = SPH
Response is new threshold settings for SPL
(ppsee) and SPH
(PPSEE).
Y = 1 (Open Collector)
Y = 2 (Relay)
W8x
x
Set sensor voltage control, where x is:
1 = menu control
2 = pin control
W10x
x
Set high voltage state, where x is:
0 = off
1 = on
Response is new high voltage setting.