45
*stb?
Returns SBR (Status Byte Register) contents.
Example: 81 means SBR=01010001
Value
Bit
Event
64
32
16
8
6
5
4
3
Service Request
Standard Event
Message Available
Questionable Data
*trg
Manually triggers the DL-2052.
7-5-10. Secondary display: CONFigure2 command
conf2:volt:dc
Configure 2
nd
display to DC Voltage.
Parameter: NR2, min, max
Example: conf2:volt:dc_1 (DC Voltage, 1V range)
conf2:volt:ac
Configure 2
nd
display to AC Voltage.
Parameter: NR2, min, max
Example: conf2:volt:ac_1 (AC Voltage, 1V range)
conf2:curr:dc
Configure 2
nd
display to DC Current.
Parameter: NR2, min, max
Example: conf2:curr:dc_10e-3 (DC Current, 10mA range)
conf2:curr:ac
Configure 2
nd
display to AC Current.
Parameter: NR2, min, max
Example: conf2:curr:ac_10e-3 (AC Current, 10mA range)
conf2:res
Configure 2
nd
display to 2W Resistance.
Parameter: NR2, min, max
Example: conf2:res_10e2 (2W Resistance, 1k
Ω range)
conf2:fres
Configure 2
nd
display to 4W Resistance.
Parameter: NR2, min, max
Example: conf2:fres_10e2 (Resistance, 1k
Ω range)
conf2:freq
Configure 2
nd
display to Frequency.
conf2:per
Configure 2
nd
display to Period.
conf2:temp
Configure 2
nd
display to Temperature.
conf2:off
Turn off the dual display mode (2
nd
display is off)
conf2:stat:func?
Returns function of 2
nd
display.
Parameter: 1 (DCV), 2 (ACV), 3 (DCA-10A), 4 (ACA-10A), 5 (DCA-mA),
6 (ACA-mA), 7 (2WR), 8 (Freq), 9 (TempC), 10 (AC+DCA-10A), 11 (AC+DCV),
12 (AC+DCA-mA), 13 (Diode), 14 (Period), 15 (TempF), 16 (4WR), 17 (Cont.)
conf2:stat:rang?
Returns range of 2
nd
display. Parameter:
DCV: 1 (100mV), 2 (1V), 3(10V), 4 (100V), 5 (1000V)
ACV: 1 (100mV), 2 (1V), 3(10V), 4(100V), 5(750V)
AC+DCV: 1 (100mV), 2 (1V), 3(10V), 4 (100V), 5 (1000V)
DCA, ACA, AC+DCA: 1(10mA), 2(100mA), 3(1A)
2WR, 4WR: 1(100Ω), 2(1kΩ), 3(10kΩ), 4(100kΩ), 5(1MΩ), 6(10MΩ), 7(100MΩ)
DCA, ACA, AC+DCA (10A range): 1 (one range)
Freq, TempC, TempF, Diode, Period, Cont.: 1 (one range)
conf2:auto
Set 2
nd
display to Auto range.
Parameter: 0 (disable auto range), 1 (enable auto range)
conf2:auto?
Return 2
nd
display Auto range status.
Parameter: 0 (disable auto range), 1 (enable auto range)