Mini-Seis III Pro™ Operating Manual
Page 39
CLI Commands
There are numerous commands that can be used to setup and control Pro operation. It is
best to use these commands with the Command Prompt in Seismograph Data Analysis. All
commands persist through a power cycle.
Command Description
Parameters
AU enable
Get or set whether the air channels are used. If enabled is 1, air
channels are included. If enabled is 0, air channels are
excluded. If passed with no parameter, returns the current
value.
enable:
0 = Air is disabled
1 = Air is enabled
AW used
Get or set whether A-weighting is used. If the A-weighting
board is not installed, the result will always be 0. If passed with
no parameter, returns the current value. If the A weight board is
installed and enabled the return value will be 2.
used:
0 = Not used
1 = Use A-weighting
BV
Get the last raw battery voltage in the format #.##.
None
CHEN
channels
Enables 8 channels (channels = 2), four channels (channels = 1),
or none (channels = 0) and configures enabled channels as
follows:
Channels 1-3: Seismic
Channel 4: Air
Channels 5-7: Seismic
Channel 8: Air
The commands SSP 0, SSP 1, or SSP 2 should be used to
configure the range, units, and min.
channels:
0 = All disabled
1 = Channels 1-4 enabled,
channels 5-8 disabled
2 = All enabled
CHG
Returns the external charging voltage.
None
CL
Clears the system log.
None.
CO text
Get or set the text of the comment preset. Text is all characters
following the space following CO (up to 126 characters, so
comments may include spaces and punctuation in addition to
numbers and letters. The line feed character (‘\n’ / 0x0A) is
treated as a carriage return. If passed with no parameters,
returns the current text for the comment.
text:
Any combination of up to
126 ASCII characters from
0x20 (space) to 0x7E
(tilde), inclusive, as well
as 0x0A (line feed).
COMP
enabled
Turn compression on/off when downloading.
Enabled:
0 = Disable compression
1 = Enable compression
CS enable
Enable or disable comment inclusion with subsequent records.
Note that comment text is set using the CO command. If sent
with no parameters, returns the current enable value.
enable:
0: Do not include the
comment preset on
subsequent records.
1: Include the comment
preset with records
collected following
enabling comments.
DU seconds
Get or set the waveform duration in seconds.
seconds:
1 to 120
EM
Erase memory
none
ER mode
[HHMM]
Set automatic reporting mode and parameters. Note that HH and
MM must be 0-padded.
If passed with no parameters and the current mode is 0 or 1,
returns only the value of the current mode.
If passed with no parameters and the current mode is 2 or 3,
returns the current values of all parameters.
mode:
0 = Disable automatic
reporting
1 = Triggered mode
2 = Fixed mode
3 = Interval mode
HH: 00–23 (required for
fixed and interval modes)