Mini-Seis III Pro™ Operating Manual
Page 41
Command Description
Parameters
RT rate
Get or set the sample rate for the current mode. If passed with
no parameters, returns the current value.
rate:
1024
2048
4096
8192
16384
65536 (Single channel
waveform only)
131072 (Single channel
waveform only)
SI rate
Set the baud rate of the serial link (only RS232 Console). If
passed with no parameters or an invalid parameter, returns the
current value. Using this command from the USB interface has
no effect.
rate:
1200
2400
4800
9600
19200*
38400
57600
115200
230400
* Default
SM mode
Get or set the current operating mode. If passed with no
parameters, returns the current value.
mode:
ID: Idle (maintains current
mode but disables data
collection)
HI: Histogram
HW: Histogram/Waveform
MA: Manual
WF: Waveform
SP seconds
Get or set histogram sample period in seconds. Period may be
1, 10, 20, 30, 40, 50, 60, or 900.
period:
1
10
20
30
40
50
60
900
SSP sensor
range unit
min
Get or set the sensor parameters for sensor-type, which can be
0 (air), 1 (seismic), or 2 (A-weighted air). If passed with no
parameters, returns an error. If passed with only sensor-type,
returns the current parameters for that type in the format:
<sensor-type as integer> <range as float> <unit as integer>
[<min as float>]
Note that min is only sent if the sensor type is 2.
Sensor-type 0 parameters affect channel 4, and channel 8 if
enabled.
Sensor-type 1 parameters affect channels 1-3, and 5-7 if
enabled.
Sensor-type 2 parameters affect channel 4 only. If sensor type 2
is selected, then channel 8 (if enabled) uses the SSP 0 settings.
Sensor is the type of
sensor. Valid values are:
0: Air
1: Seismic
2: A-Weighted Air
Range is the maximum
reading (in units) for the
channel.
Unit is the unit-type and is
sensor-type specific.
For air:
0: dBL
1: mb
2: Pa
3: PSI
For seismic:
0: IPS
1: MMPS