Operating Instructions - CPA2501
57
Air Data Test Indicator
CPA2501
:HEIGht?
Returns the head pressure height in cm
:MODE?
Returns “OFF”, “GAS”, or “LIQUID”
:MODE OFF | GAS | LIQUID
Sets the head pressure mode
:MEDium<n>
Sets the medium density
:MEDium?
Returns medium
:ACTive <C>
Sets the active channel
ACTive?
Returns the active channel
SYSTem
:DATE <i,i,i>
Not used, kept for backwards compatibility
:DATE?
Not used, does not cause an error, does not return a response
:TIME <i,i,i>
Not used, kept for backwards compatibility
:TIME?
Not used, does not cause an error, does not return a response
:ERRor[:NEXT]?
Returns error code, description
:KLOCk ON | OFF | 1 | 0
Sets the keylock state
:PRESet
Load known state values
:SAVe
No function (not needed)
:VERSion?
Returns SCPI version 1994.0
TEST
:ELECtronic?
Returns “OK”
UNIT
:[PRESsure] bar | mbar | Pa | psi
Sets the pressure units
:[PRESsure]?
Returns the pressure units
:NAME<n>?
Returns the units string for units code <n>
:FACTor <n>?
Returns the units conversion for units code <n>
:INDEX <n>
Sets the index number.
:INDEX?
Returns the index number.
index unit
0 bar
1 mbar
2 Pa
3 psi
4 atm
5 kp/cm2
6 lbf/ft2
7 kPa
8 cmH2O(4°C)
9 inH2O(4°C)
10 inH2O(60°F)
11 ftH2O(4°C)
12 µmHg(0°C)
13 mmHg(0°C)
14 cmHg(4°C)
15 inHg(0°C)
16 inHg(60°F)
17 - -
18 user
19 user
20 user
Note:
Except for the engineering units selection, the numeric suffix selects the applicable channel/transducer:
1 = Channel A
2 = Channel B
3 = Channel C
4 = Channel D
5 = baro (if installed)
This numeric suffix always defaults to 1 and is designated by [C] (channel)