44
0004-000067
Status Register Variables
The following table lists the PVs associated with the various instrument status registers.
For the Digitizer Test Register PVs in the following table,
N
is substituted with either 1 or
2 (for 4 channel instruments only) (for example: getStatusTestDig1Enable).
Process Variable
Type
Values
Description
Status mbbiDirect
00
16
to FF
16
Returns the latched instrument status:
1 = latched event
Bit 0 = unused
Bit 1 = unused
Bit 2 = error log not empty
Bit 3 = questionable event
Bit 4 = message available
Bit 5 = standard event
Bit 6 = master summary status
Bit 7 = operation event
setStatusEnable
getStatusEnable
mbboDirect
mbbiDirect
00
16
to FF
16
Status register bit enable mask
Bit 0 = unused
Bit 1 = unused
Bit 2 = error log not empty
Bit 3 = questionable event
Bit 4 = message available
Bit 5 = standard event
Bit 6 = master summary status
Bit 7 = operation event
StatusStandard mbbiDirect
00
16
to FF
16
Returns the latched standard register
events:
1 = latched event
Bit 0 = operation complete
Bit 1 = request control
Bit 2 = query error
Bit 3 = device dependent error
Bit 4 = execution error
Bit 5 = command error
Bit 6 = user request
Bit 7 = power on
setStatusStandardEnable
getStatusStandardEnable
mbboDirect
mbbiDirect
00
16
to FF
16
Standard register bit enable mask
Bit 0 = operation complete
Bit 1 = request control
Bit 2 = query error
Bit 3 = device dependent error
Bit 4 = execution error
Bit 5 = command error
Bit 6 = user request
Bit 7 = power on