ENDEVCO 6634C
INSTRUCTION MANUAL
IM6634C
Page 4-5
To determine the current output mode setting, send the following command to the 6634C (after selecting the
specific 6634C unit as Listener/Talker using the InnXXX; command):
M?XXXX ;
Request current output mode
Unit responds with:
MnXXXX "ACK"
where n is the number corresponding to byte 1 of the input type
command, i.e.:
M0XXXX "ACK" = acceleration output mode
M1XXXX "ACK" = velocity output mode
M2XXXX "ACK" = displacement output mode
Different Output engineering units can be selected for each input type (PE, Velocity Coil, RCC Input). The new
output choices are stored as the defaults for each input type in non-volatile memory.
B.
DC OUT
To remotely set the DC Output units (RMS, average, or peak) send one of the following commands to the 6634C:
D0XXXX ;
Set DC output mode to RMS
D1XXXX ;
Set DC output mode to average
D2XXXX ;
Set DC output mode to peak ("pk-pk" for displacement)
If new DC output mode is successfully setup, no response will be provided by the 6634C. If a problem is
encountered, and the unit is in the Listener/Talker mode, the "setup is not allowed" error message E1XXXX
"BEL", will be provided by the 6634C.
To determine the current DC output mode setting, send the following command to the 6634C (after selecting the
specific 6634C unit as Listener/Talker using the InnXXX; command):
D?XXXX ;
Request current DC output mode
Unit responds with:
DnXXXX "ACK"
where n is the number corresponding to byte 1 of the input type
command, i.e.:
D0XXXX "ACK" = RMS DC output mode
D1XXXX "ACK" = average DC output mode
D2XXXX "ACK" = peak DC output mode
("pk-pk" for displacement)
Different DC Output mode units can be selected for each input type (PE, Velocity Coil, RCC Input). The new
output choices are stored as the defaults for each input type in non-volatile memory.
C.
FULL SCALE
To remotely set the Full Scale output, send the following command to the 6634C:
Onnnnn ; Set full scale output to nnnnn, where nnnnn is in floating point decimal format. Its value is equal to
0.(byte0)(byte1)(byte2)(byte3) E (byte4).