U2829 Series Operation Manual Chapter 7 Command Reference
114
<value> is the Cx nominal value in NR1, NR2 or NR3 data format followed by H.
<low limit> is the low limit in NR1, NR2 or NR3 data format.
<high limit> is the high limit in NR1, NR2 or NR3 data format.
NOTE: The low limit should be smaller than the high limit, or error
information will be reported.
For example: WrtCmd (“TRAN:Cx:LIMit 0.01H, -0.01, 0.01”)
Query syntax: TRAN:Cx:LIMit?
Return format: <NR3>, <NR3><NR3><NL^END>
The TRAN:DCR:STAT command is used to set the DCR test parameter to be ON or
OFF. The TRAN:DCR:STAT? query returns the current state of the DCR test
parameter.
Command syntax:
ON
TRAN:DCR:STAT OFF
1
0
Where,
1 (decimal 49) is equal to ON.
0 (decimal 48) is equal to OFF.
For example: WrtCmd (“TRAN:DCR:STAT:ON”)
Query syntax: TRAN:DCR:STAT?
Return format: <NR1><NL^END>
The TRAN:DCR:LIMit command is used to set the DCR nominal value, high and
low limit. The TRAN:DCR:LIMit? query returns the current DCR nominal value,
high and low limits.
Command syntax: TRAN:DCR:LIMit <value>,<low limit>,<high limit>
Where,
<value> is the DCR nominal value in NR1, NR2 or NR3 data format followed by
Ω
.
<low limit> is the DCR low limit in NR1, NR2 or NR3 data format.
<high limit> is the DCR high limit in NR1, NR2 or NR3 data format
NOTE: The low limit should be smaller than the high limit, or error
information will be reported.
For example: WrtCmd(“TRAN:DCR:LIMIT 50
Ω
, -0.01, 0.01”)
Query syntax: TRAN:DCR:LIMI?
Return format: <NR3>,<NR3><NR3><NL^END>
The TRAN:MODE is used to set the transformer working mode. The TRAN:MODE?
query returns the current working mode of the transformer.
Command syntax: