14-8
SCPI Reference Tables
Model 6487 Reference Manual
[:SENSe[1]]
[:CURRent[:DC]]
:RANGe
Configure measurement range:
√
[:UPPer] <NRf>
Select range: -0.021 to 0.021 (amps).
2.1e-4
√
[:UPPer]?
Query range value.
√
:AUTO <b>
Enable or disable autorange.
(see Note)
√
:AUTO?
Query state of autorange.
√
:ULIMit <NRf>
Select autorange upper limit: -0.021 to
0.021 (amps).
2.1e-2
:ULIMit?
Query upper limit for autorange.
:LLIMit <NRf>
Select autorange lower limit: -0.021 to
0.021 (amps).
2.1e-9
:LLIMit?
Query lower limit for autorange.
:AVERage
Path to control the Digital Filter:
:TCONtrol <name>
Select filter control: MOVing or REPeat.
MOV
:TCONtrol?
Query filter control.
:COUNt <n>
Specify filter count: 2 to 100.
10
:COUNt?
Query filter count.
[:STATe] <b>
Enable or disable digital filter.
OFF
[:STATe]?
Query state of digital filter.
:MEDian
Path to control median filter:
:RANK <NRf>
Specify “n” for rank: 1 to 5 (rank = 2n+1).
1
:RANK?
Query rank.
[:STATe] <b>
Enable or disable median filter.
OFF
[:STATe]?
Query state of median filter.
:DAMPing
Path to control analog filter damping:
[:STATe] <b>
Enable or disable analog filter damping.
ON
[:STATe]?
Query filter damping state.
[:SENSe[1]]
Note: *RST default is ON and SYSTem:PRESet default is OFF.
Table 14-4 (cont.)
SENSe command summary
Command
Description
Default
parameter
Ref.
Section SCPI