and repeat the above steps for the low set point. After entering both values, 
"donE" will appear on the display. Press ENTER to store the changes or 
ESCAPE to quit. In either case "SP" will appear on the display and the ENTER 
key may be used for re-entering the set point change mode or the MODE key 
for proceeding with the normal operation of the gauge. 

 

 

7 AVERAGE 

MODE 

 

This mode is used for obtaining an 
average force or torque reading over a 
specified period of time. Applications 
include measurement of peel force, 
muscle strength, frictional force and 
any other tests requiring time-
averaged readings. There are three 
user-programmable parameters 
associated with this mode: trigger force 

or torque, initial delay and average time. The programming of these parameters 
and the enabling of the Average mode are done during the instrument setup. 
Please refer to the "Configuration" section for more information. 
 
Press MODE until "A" is displayed and then CLEAR or ZERO to begin testing. 
The process of averaging begins as soon as the programmed trigger reading is 
reached and is indicated by a flashing "A". The conclusion of the test is 
indicated by an alternating "A" and the calculated value. The readings obtained 
during the initial delay will not be part of the average, but the peak value is 
stored for later recall. A new test may be started by pressing CLEAR or ZERO. 
 

GAUGE CONTROL LANGUAGE         

The indicator can be controlled by an external device through the RS-232 
channel. The following is a list of supported commands and their interpretations. 
All commands must be terminated with a Carriage Return character (hex 0D) or 
with a Carriage Return/Line Feed combination (hex 0D+0A). The indicator 
responses are always terminated  with a Carriage Return/Line Feed. 
 

Enable Average mode 

AD 

Disable Average mode 

AM 

Select Average mode (if enabled) 

ATn 

Average time. n=0.1-300.0 seconds 

DELn 

Initial delay. n=0.1-300.0 seconds 

TRFn 

Trigger force or torque. n=value (+ for compression (CW),  

 

- for tension (CCW) 

 
AOFFn Auto-shutoff. 

n=0,1,5,10,20,30 minutes. 0=always on 

AOUTn 

Auto-transmit every nth reading. n=0,1,2,4,8,16,32,64,128. 

 0=disabled 
 
 

DELAY

AVERAGE TIME

TIME

FORCE

PEAK

AVG

TRIGGER

FORCE

Summary of Contents for BGI

Page 1: ...DIGITAL FORCE TORQUE INDICATOR Model BGI User s Guide...

Page 2: ...ed For additional information or answers to your application questions contact us and our technical support and engineering teams will be eager to help you Thank you again for your purchase and happy...

Page 3: ...in parentheses For a detailed description of the secondary functions see Section 3 POWER ENTER Turns power on and off UNITS Selects units of measurement ZERO Zeroes any tare value up to the full capac...

Page 4: ...has several features with programmable options allowing many user specified choices In order to access the configuration menu perform the following 1 Turn off the indicator 2 Press and hold MODE 3 Tu...

Page 5: ...Set point outputs enabled bcd d Mitutoyo BCD output disabled bcd E Mitutoyo BCD output enabled nPOL Mitutoyo readings without polarity absolute value POL Mitutoyo readings with polarity positive for...

Page 6: ...gs A E Average mode enabled A d Average mode disabled dEL Initial delay prompt At Average time prompt trF Trigger force torque value prompt CAL Calibration See Section 10 4 FILTERS For maximum flexibi...

Page 7: ...trigger signal returns to its original state Please refer to the diagram for connection details and to Section 3 for configuration information 6 SET POINTS This feature is useful for tolerance checkin...

Page 8: ...The process of averaging begins as soon as the programmed trigger reading is reached and is indicated by a flashing A The conclusion of the test is indicated by an alternating A and the calculated val...

Page 9: ...e PKCW Peak Clockwise mode PKCCW Peak Counter Clockwise mode CLR Clear peaks start a new average or external trigger test Z Zero display and perform the CLR function Request the displayed reading C Re...

Page 10: ...ngs and features using the same abbreviations as the commands to set them Any detected errors are reported back by means of the following error codes 10 Illegal command 11 Not applicable e g SPHn comm...

Page 11: ...transmitted if DATA is pressed during the average computation while in the Average mode or during the input scanning in the External trigger mode Mitutoyo BCD This output is useful for connection to...

Page 12: ...eration In some cases the display will show nnnn or uuuu to indicate excessive or insufficient calibration weight This can be caused by incorrect weights tare weight of over 10 of the full capacity or...

Page 13: ...cts in workmanship and materials Mark 10 Corporation will at its option repair or replace or refund the purchased price of goods found to be defective This remedy shall be the buyer s sole and exclusi...

Page 14: ...standard line of products we can provide modifications and custom designs for OEM applications Our engineering team is eager to satisfy any special requirements Please contact us for further informati...

Reviews: