National Instruments Corporation
21
NI 5122/5124 Calibration Procedure
3.
Call
niScope_ConfigureHorizontalTiming
(niScope Configure
Horizontal Timing VI) with the following parameters:
•
minSampleRate
:
10,000,000
•
minNumPts
:
100,000
•
refPosition
:
50.0
•
numRecords
:
1
•
enforceRealtime
:
NISCOPE_VAL_TRUE
4.
Call
niScope_Commit
(niScope Commit VI) with the following
parameter:
•
vi
: The instrument handle you obtained from
niScope_init
5.
Connect the calibrator test head directly to the channel 0 input of the
digitizer and output 0.75 V with a 50
Ω
load impedance.
6.
Wait 2,500 ms for the impedance matching of the calibrator to settle.
7.
Call
niScope_InitiateAcquisition
(niScope Initiate
Acquisition VI) with the following parameter:
•
vi
: The instrument handle you obtained from
niScope_init
8.
Call
niScope_FetchMeasurement
(niScope Fetch
Measurement VI) with the following parameters:
•
channelList
:
"0"
•
timeout
:
1.0
•
scalarMeasFunction
:
NISCOPE_VAL_VOLTAGE_AVERAGE
9.
Output –0.75 V with the calibrator.
10. Call
niScope_ConfigureVertical
(niScope Configure
Vertical VI) with the following parameters:
•
channelList
:
"0"
•
range
:
2.0
•
offset
: –
0.75
•
coupling
:
NISCOPE_VAL_DC
•
probeAttenuation
:
1.0
•
enabled
:
NISCOPE_VAL_TRUE
11. Call
niScope_Commit
(niScope Commit VI) with the following
parameter:
•
vi
: The instrument handle you obtained from
niScope_init
12. Wait 750 ms for the output of the calibrator to settle.
13. Call
niScope_InitiateAcquisition
(niScope Initiate
Acquisition VI) with the following parameter:
•
vi
: The instrument handle you obtained from
niScope_init