National Instruments Corporation
5
NI 5122/5124 Calibration Procedure
Documentation
You may find the following documentation helpful as you write your
calibration procedure:
•
NI High-Speed Digitizers Getting Started Guide
•
NI High-Speed Digitizers Help
•
NI PXI/PCI-5122 Specifications
or
NI PXI/PCI-5124 Specifications
•
NI-SCOPE Function Reference Help
or
NI-SCOPE VI Reference Help
These documents are installed with NI-SCOPE. You can also download the
latest versions from the NI Web site at
ni.com/manuals
.
Self-Calibration Procedures
The NI 5122/5124 includes precise internal circuits and references used
during self-calibration to adjust for time and temperature drift. The
digitizer gain, offset, flatness, and trigger timing are all corrected in
self-calibration. Measuring the accuracy of these internal calibration
circuits with another instrument during external calibration provides
traceability for the verification procedure. Absolute accuracy is ensured by
compensating for any error measured in the internal references.
Note
Self-calibrate the digitizer before you perform verification. NI-SCOPE includes
self-calibration example programs for LabVIEW, CVI, and Microsoft Visual C.
You can initiate self-calibration using Measurement & Automation
Explorer (MAX), using the Scope Soft Front Panel (SFP), or
programmatically using NI-SCOPE.
IVI\Include\niScopeCal.h
Calibration header file, which you must include
in any C program accessing calibration
functions. This file automatically includes
niScope.h
, which defines the rest of the
NI-SCOPE interface
IVI\Drivers\niScope\Examples\
Directory of NI-SCOPE examples for CVI, C,
Visual C++, and Visual Basic
Table 1.
Calibration File Locations after Installing NI-SCOPE 2.7 or Later (Continued)
File Name and Location
Description