National Instruments Corporation
21
NI PXI-4070 Calibration Procedure
14. Output 500 mV at 30 Hz on the calibrator.
15. Call
niDMM_ConfigureMeasurement
with the following
parameters:
•
Function
=
NIDMM_VAL_AC_VOLTS_DCCOUPLED
•
Range
=
0.5
•
Resolution
= 500
e–9
16. Call
niDMM_Read
. Verify that this measurement falls between the
limits listed in Table 12.
17. Refer to Table 3 for the appropriate calibrator outputs and parameter
values as you complete the following steps:
a.
On the calibrator, output the value listed under Calibrator Output
in Table 3 for the current iteration.
b.
Call
niDMM_ConfigureMeasurement
with
Mode
set to
NIDMM_VAL_AC_VOLTS
and the remaining parameters as shown
in Table 3 for the current iteration.
c.
Call
niDMM_Read
. Verify that this measurement falls between the
limits listed in Table 12.
d.
Call
niDMM_ConfigureMeasurement
again, changing
Mode
to
NIDMM_VAL_AC_VOLTS_DCCOUPLED
.
e.
Call
niDMM_Read
. Verify that this measurement falls between the
limits listed in Table 12.
5
50 mV
50 kHz
NIDMM_VAL_AC_VOLTS
0.05
50e–9
50 mV
50 kHz
NIDMM_VAL_AC_VOLTS
_DCCOUPLED
0.05
50e–9
6
50 mV
100 kHz
NIDMM_VAL_AC_VOLTS
0.05
50e–9
50 mV
100 kHz
NIDMM_VAL_AC_VOLTS
_DCCOUPLED
0.05
50e–9
7
50 mV
300 kHz
NIDMM_VAL_AC_VOLTS
0.05
50e–9
50 mV
300 kHz
NIDMM_VAL_AC_VOLTS
_DCCOUPLED
0.05
50e–9
Table 2.
niDMM_ConfigureMeasurement Parameters (Continued)
Iteration
Calibrator Output
niDMM_ConfigureMeasurement Parameters
Amplitude
Frequency
Function
Range
Resolution