91
4004
EraseExt
w/
r
0...3
0
Clears minimum and maximum values with time
and date of occurrence on the main input
Value
Description
0
without changes
1
erasing minimum value
2
erasing maximum value
3
erasing minimum and maximum value
4005
No.OfErr
w/
r
0...10
2
Number of acceptable errors in modbus RS-485
answers when transducers interface works
in Master mode
4006
Math Fun
w/
r
0...5
0
Value
Description
0
Mathematical functions on main input
switched off
1
Square of measured value
2
Square of measured value
3
Inverse of measured value
4
Inverse square of measured value
5
Inverse square root of measured value
4007
Median
w/
r
1...50
3
Number of median filter samples
4008...
...4016
w/
r
RESERVED
4017
w/
r
0...1
0
Erase transducer status registers
Value
Description
0
without changes
1
erasing status register
4018
Dec.P 2
w/
r
0...4
0
Minimum decimal point of the second displayed
value (Value displayed on the lower line of LCD)
Value
Description
0
0.0000
1
00.000
2
000.00
3
0000.0
4
00000