22 

 

[F2 PL 4] 

Device locate: move to position 4 and reply when done.  (Device should be 

 

  

initialized prior to using this command for the first time.) 

......[F2 DL 4]  D

evice is now in position 4. 

[F2 DD 400] 

Set speed to 400 (acceptable range 100 – 900, with a default of 500). 

 
 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

13.  Reference Cuvette 

 

These commands will have an effect only for systems with two independently‐controlled sample 
holders. 

 
[R1 ...] 
 

To control and monitor the temperature and status of the reference cuvette using a Dual 
Temperature Controller, use any commands in classes 3‐8, 10 and 11, substituting 

R1

 for 

F1

.  There 

are no corresponding [R1 …] commands for command classes 1, 2, 9 and 12. 
 
If you wish to ramp the temperature of the reference and sample cuvettes together, please note the 
linking command, 

[F1 TL +]

, in command class 10. 

 
 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

15.  Control of Automatic Reports 

 

The TC 1 can be controlled manually using the buttons and the display on the front.  When a 
computer program is in use, the TC 1 will automatically send reports to the program whenever a 
manual change has been made.  The commands in this section are intended to allow the program 
to prevent the TC 1 from sending those automatic reports.

 

 

[F1 XX R(+/-)] 

Stop (R‐) or start (R+) reporting all manual changes, including those 

starting with [R1 and [F2. 

 

 

 

By replacing the XX with individual command codes, a program can 

selectively stop or start reporting of manual changes. 

 

[F1 SS R-] 

Stop reporting manual changes for the sample stirrer. 

[F1 TC R-] 

Stop reporting manual changes for the sample temperature control. 

[F1 TT R-] 

Stop reporting manual changes for the sample target temperature. 

[F1 PR R-] 

Stop reporting manual changes for the external probe connection. 

[F1 RR R-] 

Stop reporting manual changes for the sample ramp rate. 

 

[F2 PL R-] 

Stop reporting manual changes for the sample position. 

 
[R1 SS R-] 

Stop reporting manual changes for the reference stirrer. 

[R1 TC R-] 

Stop reporting manual changes for the reference temperature control. 

Summary of Contents for Turret 6/AP

Page 1: ...el 509 624 9290 Fax 509 624 9488 E mail quantum qnw com Web www qnw com Turret 6 AP 6 Position Temperature Controlled Cuvette Turret for the Applied Photophysics Chirascan Manual Product Overview Janu...

Page 2: ...Turret 6 8 A General Description 8 B Cuvettes 9 C Cuvette z Height 9 D Cooling Water for the Peltier 9 E Dry Gas Purging 10 F External Temperature Probe 10 G External Computer Control 10 H Manual Rese...

Page 3: ...mputer control of the TC 1 Temperature Controller With this software the temperature of the sample holder and a temperature sensed by an external probe may be plotted vs time Simple text scripts may b...

Page 4: ...t start automatically locate the SETUP exe file in the root directory and run it The installation process starts with a small black window that is shown during installation of the drivers needed to co...

Page 5: ...Press SET to initiate temperature control The temperature will rise and stabilize indicated by the change in the green LED light on the front panel from flashing to steady To control stirring go to t...

Page 6: ...ess SET to retain this new Target and initiate temperature control The green LED light will begin flashing slowly as the device seeks the targeted temperature and the page window will show the Current...

Page 7: ...is 0 01 C minute If needed much slower ramps may be set through software commands Discontinue Ramping Press SET while the Ramp is on shown by nonzero Current min to turn the Ramp off The page will upd...

Page 8: ...ve rotating magnet Dry gas enters a 1 16 inch hose barb on the front of the instrument and is carried through a manifold in the floor of the turret to small holes on the bottom of the 18 cuvette windo...

Page 9: ...polarity is reversed and it transfers heat from the heat exchanger to the tower When cooling it is particularly important to transfer this heat away from the Peltier This is accomplished with flowing...

Page 10: ...1 Temperature Controller This jack will accept the plug on a standard Series 400 or Series 500 thermistor probe When a probe is plugged into the jack the probe temperature is presented on the display...

Page 11: ...Controller The Turret 6 will home to a starting position then move to the cuvette position specified by the controller Figure 7 Cuvette position 1 shown squarely in the spectrometer light beam Cuvett...

Page 12: ...out of range warnings loose cable check cable The temperature controller is not receiving reasonable responses from either the cell tower or heat exchanger sensors Since it is very unlikely for both...

Page 13: ...ature range 20 to 110 C temperature precision 0 02 C cuvette size outside dimensions 12 5 x 12 5 mm minimum cuvette height 30 mm cuvette z height 15 mm optical port dimensions 12 mm high x 10 mm wide...

Page 14: ...om the set temperature of temperature readings returned from the temperature controlled Turret 6 for set temperatures between 15 C and 110 C Thus it is a measure of the ability of the cuvette holder t...

Page 15: ...roducibility is a measure of the ability of the temperature to return to an original value see Figure 1 It accounts for differences depending on the direction of temperature change and differences fro...

Page 16: ...re shown the time required to get within 1 C the time to get within the lock on temperature of 0 05 C from the target and the final time by which the TC 1 indicates the temperature is stable Temperatu...

Page 17: ...e will be established through a USB connection between the computer and the controller In this case the controller includes electronics which convert the USB connection to a serial communications port...

Page 18: ...r speed is 2500 rpm F1 LS What is the lowest stirrer speed F1 MS 300 The lowest stirrer speed is 300 rpm F1 SS S 1000 Set stirrer speed to 1000 rpm F1 SS S 0 Turn stirrer off and set the speed to zero...

Page 19: ...top automatic reports of instrument status F1 IS R The controller has been powered off and back on again 7 Current Sample Temperature F1 CT What is the current temperature of the holder F1 CT 22 84 Th...

Page 20: ...n F1 PA Start automatic reporting of probe temperature every temperature increment set by the command above F1 PT 30 5 The current probe temperature is 30 5 degrees F1 PA Stop automatic reporting of p...

Page 21: ...l send a response of the form F1 TT as a notification of the end of the ramp This response can be blocked at any time by previously sending the command F1 TT see command class 5 11 Heat Exchanger Temp...

Page 22: ...buttons and the display on the front When a computer program is in use the TC 1 will automatically send reports to the program whenever a manual change has been made The commands in this section are i...

Page 23: ...23 R1 TT R Stop reporting manual changes for the reference target temperature R1 RR R Stop reporting manual changes for the reference ramp rate...

Reviews: