22
[F2 PL 4]
Device locate: move to position 4 and reply when done. (Device should be
initialized prior to using this command for the first time.)
......[F2 DL 4] D
evice is now in position 4.
[F2 DD 400]
Set speed to 400 (acceptable range 100 – 900, with a default of 500).
13. Reference Cuvette
These commands will have an effect only for systems with two independently‐controlled sample
holders.
[R1 ...]
To control and monitor the temperature and status of the reference cuvette using a Dual
Temperature Controller, use any commands in classes 3‐8, 10 and 11, substituting
R1
for
F1
. There
are no corresponding [R1 …] commands for command classes 1, 2, 9 and 12.
If you wish to ramp the temperature of the reference and sample cuvettes together, please note the
linking command,
[F1 TL +]
, in command class 10.
15. Control of Automatic Reports
The TC 1 can be controlled manually using the buttons and the display on the front. When a
computer program is in use, the TC 1 will automatically send reports to the program whenever a
manual change has been made. The commands in this section are intended to allow the program
to prevent the TC 1 from sending those automatic reports.
[F1 XX R(+/-)]
Stop (R‐) or start (R+) reporting all manual changes, including those
starting with [R1 and [F2.
By replacing the XX with individual command codes, a program can
selectively stop or start reporting of manual changes.
[F1 SS R-]
Stop reporting manual changes for the sample stirrer.
[F1 TC R-]
Stop reporting manual changes for the sample temperature control.
[F1 TT R-]
Stop reporting manual changes for the sample target temperature.
[F1 PR R-]
Stop reporting manual changes for the external probe connection.
[F1 RR R-]
Stop reporting manual changes for the sample ramp rate.
[F2 PL R-]
Stop reporting manual changes for the sample position.
[R1 SS R-]
Stop reporting manual changes for the reference stirrer.
[R1 TC R-]
Stop reporting manual changes for the reference temperature control.