Section 5: Software
Calibration Variables (LIST C)
The calibration settings control the calibration and validation of the MPS4000. Reference Scanivalve Document: MPS Field
Val if performing a single temperature validation with an SPC4000.
List Calibration Settings (LIST C)
DESCRIPTION
Returns all of the calibration settings
SYNTAX
LIST C
RETURNS
All of the calibration settings followed by the prompt.
EXAMPLE
LIST C
SET NUMPTS 0 9 0 9
SET MIN 40.000000 -5.500000 5.000000 -5.000000
SET MAX 60.000000 5.500000 55.000000 5.000000
SET DELAY 5 5
SET IPOVEN 0.0.0.0 0
SET IPCAL 0.0.0.0 0 1
SET CALAVG 1 1
SET VALZO 0
SET FCAL 3 -5.00 5.00
SET FVAL 3 -5.00 5.00
>
Set Number of Points in Conversion Table (NUMPTS)
DESCRIPTION
Sets the number calibration and validation temperature and pressure points. This setting works
with the MIN and MAX setting to the calibration and validation points.
SYNTAX
SET NUMPTS <tcal> <pcal> <tval> <pval>
Where:
<tcal> - Number of temperature points for a calibration
<pcal> - Number of pressure points for a calibration
<tval> - Number of temperature points for a validation
<pval> - Number of pressure points for a validation.
These values are not range checked.
RETURNS
Prompt when ready to accept a new command
EXAMPLE
SET NUMPTS 5 9 5 9
>
DEFAULT
5 9 5 9
RANGE
<0 to 15> <0 to 25> <0 to 15> <0 to 25>
Summary of Contents for MPS4264
Page 1: ...MPS4264 Miniature Pressure Scanner Hardware and Software Manual Software Version 2 07...
Page 2: ......
Page 3: ......
Page 6: ...INTENTIONALLY LEFT BLANK...
Page 12: ...6 MPS4264 Section 1 Specifications INTENTIONALLY LEFT BLANK...
Page 18: ...12 MPS4264 Section 2 Introduction INTENTIONALLY LEFT BLANK...
Page 32: ...26 MPS4264 Section 4 Hardware INTENTIONALLY LEFT BLANK...
Page 38: ...32 MPS4264 Section 5 Software INTENTIONALLY LEFT BLANK...
Page 81: ...75 MPS4264 Section 6 Maintenance INTENTIONALLY LEFT BLANK...
Page 86: ...80 MPS4264 Section 6 Maintenance INTENTIONALLY LEFT BLANK...
Page 87: ...81 MPS4264 Appendix INTENTIONALLY LEFT BLANK...