22
WIKA operating instructions process pressure transducer, model CPT6020
0020507001F 05/2019 EN
EN
6.1.3 MENSOR Legacy Command Set
This command and query set replicates the communications with legacy Mensor products – the 61xx, CPT61xx, 6000, CPT6010
and most of the removable sensors from Mensor pressure calibrators (PCS400M, PCS 6000, CPC6000, CPC8000). The
command set enabled and stored sets the power-on command set for the transducer. From any command set, the CMD_SET
command can select which one is used.
Command Set 1 – Mensor Legacy
P/W
Command
Command String
Return String
Description
None
?
#X?
X<sp>±Y
Basic query to return
readings. Pressure in
native units are returned
None
A<x>
#X<sp>Y
R
Sets address of sensor X
to Y (0-9,A-Z)
None
CMD_SET<Y>
#XCMD_SET<sp>Y
R
Change CPT6020 into
different command sets.
1 is Mensor Legacy. 0 is
Sensor command set
None
DC?
#XDC?
X<sp>DC<sp>Y
Returns the date of
calibration where Y is
mmddyy
Cust
DC<Y>
#XDC<sp>Y
R
Loads the date of
calibration where Y is
mmddyy
None
DT?
#XDT?
X<sp>DT<sp>NNN
Returns the temperature
a/d DAC offset
None
FL?
#XFL?
X<sp>FL<sp>NN
Returns the filter – it is the
percentage of old reading
added to new reading: 0
to 99
None
FL<X>
#XFL<sp>NN
R
Sets the filter percentage
None
FS?
#XFS?
X<sp>FS<sp>NN
Returns the % FS
uncertainty
None
ID?
#XID?
X<sp> ID <sp> AAAAAAAA,
<sp>BBBBBBBB,
<sp>CCCCCCCC,
<sp>VD.DD
Returns transducer id.
‘AA…’ is an id string,
‘BB….’ is a model string,
‘CC…’ is the serial
number string, D.DD is the
firmware version
None
M?
#XM?
X<sp>M<sp> Y
Returns the operation
mode
None
R-?
#XR-?
X<sp>R-<sp>Y
Returns range floor value
None
R+?
#XR+?
X<sp>R+<sp>Y
Returns range ceiling
value
None
SAVE
#XSAVE
R
Saves all data to
EEPROM. Ready char
is sent when save is
complete
Operation