2-72
Agilent 4155B/4156B VXIplug&play Driver User’s Guide, Edition 4
Driver Functions
hpe5250a_selectCompenFile
hpe5250a_selectCompenFile
This function specifies capacitance/conductance compensation data file used to
compensate C/G by using hpe5250a_compenC.
Syntax
ViStatus _VI_FUNC hpe5250a_selectCompenFile(ViSession vi,
ViString file_name);
Parameters
vi
Instrument handle returned from hpe5250a_init( ).
file_name
Compensation data file name. Use absolute path. If the value is
NULL string, the default data is used.
Remarks
If you change the compensation data, copy the default data shown below, and
modify the data for your measurement cable. You will need to change the data for
DATA05 and 06, and/or DATA07 and 08 corresponding to your cables. To measure
and change the compensation data, refer to Agilent E5250A
User’s Guide
. To get
the R, L, and C value, measure R, L, and C of the cable using the 4284A, and divide
them by cable length (in m). Compensation data must be the value for 1 m length.
Do not change the data format in the file.
!"##$
%&%'( )*) )+), -$
!"
$.$
/ %0
*/10
/20
,3$+4.
53$46
-$3345
27%5
5
,$354.
6$46
+,$5.45
27%
.5$354.
3$46
5,-$-45
7%1&%
.
5,,$+4.
.6$46
5.$345
7%* 8
3
5$,4.
3$46
-$45
#/90
5$,4.
3$46
-$45
#1/90
+
5$,4.
3$46
-$45
#*/90
,
553$4.
33$46
5.$45
#1/90
-
553$4.
33$46
5.$45
#*/90
6
$4.
$46
5$45
":