1-120
Chapter 1
Alphabetical Command Reference
LIM
LIM
Syntax
LIM<D|L|M|S|U><num>[DB|HZ];
Description
These commands edit a limit test segment. The limit table editing is begun with
EDITLIML;
, and a segment is brought up for editing with
SEDI<num>;
or added using
SADD;
. The segment is closed with
SDON;
, the table is closed with
EDITDONE;.
NOTE
Currently these commands can be queried by sending the command followed
by the
OUTPACTI
command, as in the following example to query the upper
limit value:
10 OUTPUT 716;”LIMU;OUTPACTI;”
Future revisions of firmware may support the standard query form (which
currently always returns a zero) for these commands.
Command
Description
Range
Query Response
LIMD
Sets the limit delta value while editing a
limit line segment.
amplitude range
1
1. For log mag:
±
500 dB. For phase:
±
500 degrees. For Smith chart and Polar:
±
500 units. For linear magnitude:
±
500 units. For SWR:
±
500 units. The scale is always positive, and has minimum values of 0.001 dB, 10e-12
degrees, 10e-15 seconds, and 10 picounits.
see “Note” below
LIML
Sets the lower limit value.
amplitude range
see “Note” below
LIMM
Sets the middle limit value.
amplitude range
see “Note” below
LIMS
Sets the limit stimulus break point.
stimulus range
2
2. For frequency or power sweeps, refer to “Preset State and Memory Allocation,” in your analyzer’s reference
guide. For CW time: 0 to 24 hours. For frequency sweep, transform on:
±
1/frequency step. For CW time sweep,
transform on:
±
1/time step.
see “Note” below
LIMU
Sets the upper limit value.
amplitude range
see “Note” below
Summary of Contents for 8719ES
Page 15: ...1 1 1 Alphabetical Command Reference ...
Page 293: ...2 1 2 Introduction to Instrument Control ...
Page 310: ...3 1 3 GPIB Programming ...
Page 334: ...4 1 4 Reading Analyzer Data ...
Page 343: ...5 1 5 Data Processing Chain ...
Page 350: ...6 1 6 Error Reporting ...
Page 364: ...7 1 7 Programming Examples ...
Page 502: ...A 1 A Preset Conditions ...
Page 517: ...B 1 B Command Listings ...