Chapter 6
6-5
Error Reporting
Status Reporting
Table 6-3 Event Status Register B: Status Bit Definitions
Bit
Name
Definition
0
Single sweep, number of
groups, or calibration step
complete
A single sweep, group, or calibration step has been completed
since the last read of the register.
1
Service routine waiting or
done
An internal service routine has completed operation, or is
waiting for an operator response.
2
Data entry complete
A terminator key has been pressed or a value entered over
GPIB since the last read of the register.
3
Limit failed, Channel 2
Limit test failed on Channel 2.
4
Limit failed, Channel 1
Limit test failed on Channel 1.
5
Search failed, Channel 2
A marker search was executed on Channel 2, but the target
value was not found.
6
Search failed, Channel 1
A marker search was executed on Channel 1, but the target
value was not found.
7
Copy Complete
A copy has been completed since the last read of the register.