This query returns one of the following values:
Return
Value
Meaning
0
Calibration is complete.
New constants are in RAM.
1
More measurements are needed.
5
Calibration aborted.
•
A return value of 5 (calibration aborted) is given if either:
a. The command is received when the module is in current calibration or
resistance calibration mode, or
b. The calibration in process has entered more than 50 values without a
successful calibration.
•
When the channel voltage calibration is completed, the new calibration constants
are used by the module, but not stored in non-volatile memory.
•
If a voltage calibration is interrupted before completion (that is, by another
command being received), no constants are changed.
•
Refer to Appendix D, “Voltage/Current Output Adjustment”, starting on page 157,
for additional information about this command.
•
Related Commands:
CALibration:STORe
,
CALibration
n
:CONFigure:VOLTage?
CALibration
Chapter 3
HP E1418A SCPI Command Reference 74
Summary of Contents for E1418A
Page 6: ...Notes 6 Contents HP E1418A 8 16 Channel D A Converter Module ...
Page 10: ...Notes 10 HP E1418A User s Manual ...
Page 12: ...12 HP E1418A User s Manual ...
Page 105: ...TRIGger 105 HP E1418A SCPI Command Reference Chapter 3 ...
Page 111: ...Notes HP E1418A Command Quick Reference 111 HP E1418A SCPI Command Reference Chapter 3 ...
Page 135: ...135 HP E1418A Register Based Programming Appendix B ...
Page 157: ...Notes 156 HP E1418A Error Messages Appendix C ...
Page 170: ...Notes Appendix D Voltage Current Output Adjustment 169 ...
Page 174: ...Figure E 1 8 Channel Disassembly 172 Configuration and Disassembly Appendix E ...
Page 175: ...Figure E 2 16 Channel Disassembly Appendix E Configuration and Disassembly 173 ...
Page 192: ...Notes 192 HP E1418A 8 16 Channel D A Converter Module Index ...