Command, for module
Description
:
CONF
igure
:AVER
age
˽
{1|16|64|256|512|1024}
Set the number of digital filter averaging steps.
Factory default is 64.
Example: :CONF:AVER 16
˽
:AVER
age?
Query the digital filter averaging steps.
Example: :CONF:AVER?
→
16
:KILL
{1|0}
˽
Set function kill enable (1) or kill disable (0).
Factory default is Kill Disable.
Example: :CONF:KILL 0
˽
:KILL?
Query the current value for the kill enable function
Example: :CONF:KILL?
→
0
:ADJ
ust {1|0}
˽
Set the fine adjustment function on (1) or off (0)
Factory default is Fine adjustment on.
Example: :CONF:ADJ 1
˽
:ADJ
ust
?
Query the fine adjustment state
Example: :CONF:ADJ?
→
1
:EV
ent
˽
CLEAR
Reset the Module Event Status register (8.7).
Example: :CONF:EVENT CLEAR
˽
:MASK
<Word>
˽
Set the Module Event Mask register (8.8).
Example: :CONF:EVENT:MASK 0
˽
:MASK?
Query the Module Event Mask register (8.8).
Example: :CONF:EVENT:MASK?
→
0
:CHANMASK
˽
<Word>
Set the Module Event Channel Mask register (8.11).
Example: :CONF:EVENT:CHANMASK 255
˽
:CHANMASK?
Query the Module Event Channel Mask register (8.11).
Example: :CONF:EVENT:CHANMASK?
→
255
:CAN
:ADDR
ess {0...63}
˽
Set the modules CAN bus address (0…63).
Can only be set in configuration mode.
Example: :CONF:CAN:ADDR
˽
23
:ADDR
ess
?
Query the modules CAN bus address
Example: :CONF:CAN:ADDR?
→
23
:BIT
rate {125000|250000}
˽
Set the CAN bus bit rate to 125 kBit/s or 250 kBit/s.
Can only be set in configuration mode.
Example: :CONF:CAN:BITRATE
˽
250000
:BIT
rate
?
Query the modules CAN bus bit rate
Example: :CONF:CAN:BITRATE?
→
250000
:SERIAL
:ECHO?
Query if there is serial echo is enabled (1) or disabled (0).
Example: :CONF:SERIAL:ECHO?
→
1
:ECHO
{1|0}
˽
1: The device echos all characters received on the serial interface
(factory default)
0: The device does not echo received characters on the serial
interface. Be careful when switching off the echo as there is no
other possibility to synchronize the HV device with the computer
(no hardware/software handshake).
This mode is only available for compatibility reasons and without
support.
SCPI protocol common instruction set | Last changed on: 2020-10-07 |
www.iseg-hv.com
29/61