Model DMM7510 7½ Digit Graphical Sampling Multimeter Reference Manual
Section 6: SCPI command reference
DMM7510-901-01 Rev. B / May 2015
6-183
:TRIGger:BLOCk:BRANch:COUNter:COUNt?
This command returns the count that the trigger model is on.
Type
Affected by
Where saved
Default value
Query only
Recall settings
Instrument reset
Power cycle
Not applicable
Not applicable
Usage
:TRIGger:BLOCk:BRANch:COUNter:COUNt? <blockNumber>
<blockNumber>
The sequence of the block in the trigger model
Details
The query returns the number of times the trigger model has looped. The counter is defined by
:TRIGger:BLOCk:BRANch:COUNter
.
Example
TRIG:LOAD "Empty"
TRIG:BLOC:BUFF:CLEAR 1
TRIG:BLOC:MEAS 2
TRIG:BLOC:BRAN:COUN 3, 5, 2
TRIG:BLOC:DEL:CONS 4, 1
TRIG:BLOC:BRAN:COUN 5, 3, 2
INIT
TRIG:BLOCK:BRAN:COUN:COUN?
Reset trigger model settings.
Clear
defbuffer1
at the beginning of the
trigger model.
Loop and take 5 readings.
Delay 1 s.
Loop three more times back to block 2.
At end of execution, 15 readings are stored in
defbuffer1
.
Check to see which count the trigger model has
completed.
Also see
(on page 6-182)
:TRIGger:BLOCk:BRANch:COUNter:RESet
This command creates a block in the trigger model that resets a branch counter to 0.
Type
Affected by
Where saved
Default value
Command only
Recall settings
Instrument reset
Power cycle
Save settings
Not applicable
Usage
:TRIGger:BLOCk:BRANch:COUNter:RESet <blockNumber>, <counter>
<blockNumber>
The sequence of the block in the trigger model
<counter>
The block number of the counter that is to be reset