12-64 Instrument Control Library
Series 2600 System SourceMeters Reference Manual
2600S-901-01 Rev. A / May 2006
Return to
smuX.cal.save
X = SMU channel (a or b)
Function
Stores the calibration constants in non-volatile memory.
Usage
smuX.cal.save()
Remarks
• This function will store the current set of calibration constants in non-volatile
memory. The previous calibration constants (from the default calibration set) will be
copied to the previous calibration set (
CALSET_PREVIOUS
) prior to overwriting the
default calibration set.
•
This function will be disabled until a successful call to
smuX.cal.unlock
is
made.If any of the calibration constants have been changed, this function will be
disabled unless both the calibration date and the calibration due date have been
assigned new values.
Details
(calibration).
Also see
Example
Stores calibration constants for SMU A in non-volatile memory:
smua.cal.save()
smuX.cal.state
X = SMU channel (a or b)
Attribute
Calibration state.
Usage
calstate = smuX.cal.state
Remarks
When reading this read-only attribute,
calstate
returns one of the following values:
0
smuX.CALSTATE_LOCKED
Calibration is locked.
1
smuX.CALSTATE_CALIBRATING
The calibration constants or dates have
been changed but not yet saved to
nonvolatile memory.
2
smuX.CALSTATE_UNLOCKED
Calibration is unlocked but none of the
calibration constants or dates have
changed since the last save/restore.
Details
(calibration).
Also see
,
Example
Reads calibration state for SMU A:
calstate = smua.cal.state
print(calstate)
Output:
0.00
The above output indicates that calibration is locked.