14-18 Display Operations
Series 2600 System SourceMeters Reference Manual
2600S-901-01 Rev. A / May 2006
Return to
display.waitkey
The
display.waitkey
function captures the keycode value for the next key
press:
key = display.waitkey()
After sending the
display.waitkey
function, the script will pause and wait for the
operator to press a front panel key. For example, if the
MEAS
key is pressed, the
function will return the value 86, which is the keycode for that key. The keycode
values are listed in
Example
: The following code will prompt the user to press the
EXIT
key to abort
the script, or any other key to continue it:
display.clear()
display.setcursor(1, 1)
display.settext("Press EXIT to Abort")
display.setcursor(2, 1)
display.settext("or any key to continue")
key = display.waitkey()
display.clear()
display.setcursor(1, 1)
if (key == 75) then
display.settext("Test Aborted")
exit()
else display.settext("Test Continuing")
end