6.
Call the ni5110 Ext Cal API.lvlib:DC Reference Cal Adjust VI with the following
settings:
•
actual voltage generated (V)
: the DC voltage present on the specified channel of
the PXIe-5113
7.
Repeat steps 3 through 6 until the
dc reference cal complete
indicator of the ni5110 Ext
Cal API.lvlib:DC Reference Cal Adjust VI returns
TRUE
.
8.
Disable the calibrator output.
Adjusting 50
Ω
DC Reference
Follow this procedure to adjust the DC gain and offset of the 50 Ω DC reference of the
PXIe-5113.
1.
Connect the calibrator test head to the specified channel of the PXIe-5113.
2.
Call the ni5110 Ext Cal API.lvlib:DC Reference Cal Initialize VI with the following
settings:
•
channel
: the specified channel
•
input impedance
: 50 Ω
3.
Call the ni5110 Ext Cal API.lvlib:DC Reference Cal Configure VI to obtain the DC
voltage to generate, and configure the calibrator with the following settings:
•
Output: the specified DC voltage
•
Load impedance: 50 Ω
4.
Enable the calibrator output if it is not already enabled.
5.
Wait 1 second for settling.
6.
Call the ni5110 Ext Cal API.lvlib:DC Reference Cal Adjust VI with the following
settings:
•
actual voltage generated (V)
: the DC voltage present on the specified channel of
the PXIe-5113
7.
Repeat steps 3 through 6 until the
dc reference cal complete
indicator of the ni5110 Ext
Cal API.lvlib:DC Reference Cal Adjust VI returns
TRUE
.
8.
Disable the calibrator output.
Reverification
Repeat the
section to determine the as-left status of the PXIe-5113.
Note
If any test fails reverification after performing an adjustment, verify that you
have met the test conditions before returning your PXIe-5113 to NI. Refer to the
Worldwide Support and Services
section for information about support resources or
service requests.
PXIe-5113 Calibration Procedure
|
© National Instruments
|