SCXI-1122 Calibration Procedure
6
ni.com
6.
Call
MIO_Config
to enable dithering on the E Series DAQ device.
Set the following parameters:
•
deviceNumber
—The value assigned by MAX
•
dither
—1
•
useAmux
—0
7.
Call
SCXI_Configure_Filter
to configure the filter mode of the
SCXI-1122 module. Set the following parameters:
•
assigned bySCXIchassisId
—The value assigned by MAX
•
moduleSlot
—1
•
channel
— –1
•
filterMode
—1
•
freq
—4.0
•
cutoffDivDown
—0
•
and outClkDivDown
—2
•
ActualFreq
—This is a return value
8.
Call
SCXI_Single_Chan_Setup
to configure the SCXI-1122
module for single channel measurements. Set the following
parameters:
•
SCXIchassisID
—The value assigned by MAX
•
moduleSlot
—1
•
moduleChan
—0
•
DAQdeviceNumber
—The device number assigned by MAX for
the E Series DAQ device
9.
Call
SCXI_Calibrate_Setup
to auto-zero the SCXI-1122 module.
Set the following parameters:
•
SCXIchassisID
—The value assigned by MAX
•
moduleSlot
—1
•
calOp
—2
10. Call
SCXI_Set_Gain
to set the selected gain on the SCXI-1122
module. Set the following parameters:
•
SCXIchassisID
—The value assigned by MAX
•
moduleSlot
—1
•
channel
— –1
•
gain
—The gain value you want to test
Acceptable gain values are listed in Table 3.