84
Code
Meaning
Format
PR
Read
Profile
step in
memory
PR
<
Space
><
Arg1
><
Space
><
Arg2
><
CR
>
where
Arg1
= Profile number (range 1 to 4) and
Arg2
=Step number (range 1 to 15)
Return data format:
<
Arg1
><
Space
><
Arg2
><
Space
><
Arg3
><
Space
>...<
CR
><
LF
>
where
Arg1
=Profile number (range 1 to 4),
Arg2
=Step number,
Arg3
=Temperature setpoint,
Arg4
=Agitation setpoint,
Arg5
=(Reserved, reads 0.0),
Arg6
=Step hours,
Arg7
=Step minutes,
Arg8
=UV lamp, and
Arg9
=Grow lamp
PS
Profile
Stop/
PS
<
CR
>
With no other parameters, any profile currently running will be
stopped.
Start
(stops or
starts a
program
profile)
PS
<
Space
><
Arg1
><
Space
><
Arg2
><
CR
>
where
Arg1
=Profile number (range 1 to 4), and
Arg2
=Step number (range 1 to 15).
NOTE:
If
Arg2
is not included, the profile will start in Step 1.
PW
Write
Profile
step to
memory
PW
<
Space
><
Arg1
><
Space
><
Arg2
><
Space
><
Arg3
><
Space
>
...<
CR
>
where
Arg1
=Profile number (range 1 to 4),
Arg2
=Step number (range 1 to 15),
Arg3
=Temperature setpoint,
Arg4
=Agitation setpoint,
Arg5
=(Reserved); Enter 0.0,
Arg6
=Step hours (range 0 to 99),
Arg7
=Step minutes (range 0 to 59),
Arg8
=UV lamp (0 = off, 1 = on), and
Arg9
=Grow lamp (0 = off, 1 = on)
For example, the following
PW
command:
PW
<
Space
>
25.0
<
Space
>
150
<
Space
>
0.0
<
Space
>
2
<
Space
>
30
<
Space
>
0
<
Space
>
1
<
CR
>
will set Program 1, Step 1 to the following conditions:
Temperature setpoint 25.0º
Step Time 2 hours 30 minutes
Agitation setpoint 150 rpm
UV Lamp OFF
GRO Lamp ON
Innova 43 & 43R Incubator Shakers M1320-0050
User’s Guide
Summary of Contents for Innova 43
Page 2: ...2 Innova 43 43R Incubator Shakers M1320 0050 User s Guide ...
Page 10: ...10 Innova 43 43R Incubator Shakers M1320 0050 User s Guide ...
Page 79: ...79 11 1 1 CE Declaration of Conformity New Brunswick Scientific Co Inc User s Guide ...
Page 92: ...92 Filling the 33 Wavy Arrows Icon 21 Innova 43 43R Incubator Shakers M1320 0050 User s Guide ...