MULTIDOS
RS232 Interface
Application Afterloading
20
D177.131.2/0
11 Application Afterloading
Telegrams containing parameters alter a setting. If the parameter is omitted in the telegram, the actual setting is
returned. For clarification parameters are printed in
italic
.
Example: I0044 sets the interval time, I reads the actual interval time and is answered by I0044.
11.1 Basic Settings
Telegram
Description
Response
U
x
Set/Read unit of measurement. Parameter
x
: A = Ampere, S = Gy/s, M =
Gy/min, H = Gy/h
Ux
R
x
Set/Read measuring range. Parameter
x
: L = range Low, H = range High
Rx
SET
s
Set/Read the calibration set to be applied. Parameter
s
: 1 up to 5
SETs
SETA
aa
Set/Read the status of the channels of the applied set.
aa
is a 2 digit decimal
number, whose set bits, when converted to binary representation, signal the
following status:
.0
Þ
channel 1 of the rectum probe is active
.1
Þ
channel 2 of the rectum probe is active
.2
Þ
channel 3 of the rectum probe is active
.3
Þ
channel 4 of the rectum probe is active
.4
Þ
channel 5 of the rectum probe is active
.5
Þ
channel of the bladder probe is active
.0 is the least significant bit
aa
can take values from 00 to 63.
SETAaa
11.2 Control Telegrams
Telegram
Description
Response
NEW
Reinitialize measurement
NEW
NULE
Read the channels (only active channels are considered) with erroneous results
during last zeroing. ee is a 2 digit decimal number, whose set bits, when con-
verted to binary representation, signal the following status:
.0
Þ
channel 1 of the rectum probe was out of offset limits
...
.4
Þ
channel 5 of the rectum probe was out of offset limits
.5
Þ
channel of the bladder probe was out of offset limits
.0 is the least significant bit
ee
can take values from 00 to 63.
NULee
NULL
r
Read the offset current limits for measuring range
r
Limits (Ampere) in the data format
±m.mmmE±ee;
Data output: rectum probe channels 1 to 5, bladder channel
Example for measuring range Low:
NULLL•41.70E-12;•42.25E-12;•41.40E-12;•42.10E-12;•41.80E-12;•42.05E-12;
NULLrLimit
s
NULO
r
Read the offset current
measured during last zeroing in measuring range
r
Results (Ampere) in the data format
±m.mmmE±ee
Data output: rectum probe channels 1 to 5, bladder channel
Results for deactivated channels are 0.0
Example for measuring range Low with deactivated channel 1:
NULOL••0.00E-12;••0.60E-12;••0.50E-12;•-0.10E-12;••0.70E-12;••0.55E-12;
NULLOr-
Results